Evidence map›Paper›PMID 30753624›Full record

ArticleOperative neurosurgery (Hagerstown, Md.)2019

First-in-Human Experience With Integration of a Hydrocephalus Shunt Device Within a Customized Cranial Implant.

Chad R Gordon, Amir Wolff, Gabriel F Santiago, Kenneth Liebman, Erol Veznedaroglu, Frank D Vrionis, Judy Huang, Henry Brem, Mark Luciano

Open access · hybridAbstract readCase ReportsVideo-Audio Media
In one paragraph

Article in Operative neurosurgery (Hagerstown, Md.), 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.

0numbers the graph read from it
0cells of the map it votes in
8citing papers in PubMed
2.5field-weighted citation impact, top 11% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

8 citing papers in PubMed, 34 citations in OpenAlex.

  1. Article
  2. Review
  3. Review
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors at 2 institutions in 1 country.

Chad R GordonDivision of Neuroplastic and Reconstructive Surgery, Department of Plastic and Reconstructive Surgery, Surgery, Department of Plastic and Reconstructive Surgery, Johns Hopkins University School of Medicine, The Johns Hopkins Hospital, Baltimore, Maryland.
Amir WolffDivision of Neuroplastic and Reconstructive Surgery, Department of Plastic and Reconstructive Surgery, Surgery, Department of Plastic and Reconstructive Surgery, Johns Hopkins University School of Medicine, The Johns Hopkins Hospital, Baltimore, Maryland.
Gabriel F SantiagoDivision of Neuroplastic and Reconstructive Surgery, Department of Plastic and Reconstructive Surgery, Surgery, Department of Plastic and Reconstructive Surgery, Johns Hopkins University School of Medicine, The Johns Hopkins Hospital, Baltimore, Maryland.
Kenneth LiebmanDepartment of Neurosurgery, Drexel College of Medicine, Philadelphia, Pennsylvania.
Erol VeznedarogluDepartment of Neurosurgery, Drexel College of Medicine, Philadelphia, Pennsylvania.
Frank D VrionisMarcus Neuroscience Institute, Boca Raton Regional Hospital, Boca Raton, Florida.
Judy HuangDepartment of Neurosurgery, Johns Hopkins University School of Medicine, The Johns Hopkins Hospital, Baltimore, Maryland.
Henry BremDepartment of Neurosurgery, Johns Hopkins University School of Medicine, The Johns Hopkins Hospital, Baltimore, Maryland.
Mark LucianoDepartment of Neurosurgery, Johns Hopkins University School of Medicine, The Johns Hopkins Hospital, Baltimore, Maryland.
Johns Hopkins University · USNeurosciences Institute · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundImplantable shunt devices are critical and life saving for hydrocephalus patients. However, these devices are fraught with high complication rates including scalp dehiscence, exposure, and extrusion. In fact, high shunt valve profiles are correlated with increased complications compared to those with lower profiles. As such, we sought a new method for integrating shunt valves for those challenging patients presenting with scalp-related complications.

objectiveTo safely implant and integrate a hydrocephalus shunt valve device within a customized cranial implant, in an effort to limit its high-profile nature as a main contributor to shunt failure and scalp breakdown, and at the same time, improve patient satisfaction by preventing contour deformity.

methodsA 64-yr-old male presented with an extruding hydrocephalus shunt valve and chronic, open scalp wound. The shunt valve was removed and temporary shunt externalization was performed. He received 2 wk of culture-directed antibiotics. Next, a contralateral craniectomy was performed allowing a new shunt valve system to be implanted within a low-profile, customized cranial implant. All efforts were made, at the patient's request, to decrease the high-profile nature of the shunt valve contributing to his most recent complication.

resultsFirst-in-human implantation was performed without complication. Postoperative shunt identification and programming was uncomplicated. The high-profile nature of the shunt valve was decreased by 87%. At 10 mo, the patient has experienced no complications and is extremely satisfied with his appearance.

conclusionThis first-in-human experience suggests that a high-profile hydrocephalus shunt device may be safely integrated within a customized cranial implant.

Indexed as

Computer-Aided DesignProsthesis DesignProsthesis FailureHumansHydrocephalusMaleMiddle AgedScalpVentriculoperitoneal ShuntCranial implantCranioplastyHydrocephalus shuntImplant extrusionScalp woundShunt complicationVentriculoperitoneal shunt

Identifiers

PMID30753624
PMCPMC6855953
OpenAlexW2914559454

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.