Evidence map›Paper›PMID 30735994›Full record

ArticleIEEE transactions on ultrasonics, ferroelectrics, and frequency control2019

Combining Slow Flow Techniques With Adaptive Demodulation for Improved Perfusion Ultrasound Imaging Without Contrast.

Jaime Tierney, Kristy Walsh, Helen Griffith, Jennifer Baker, Daniel B Brown, Brett Byram

Open access · greenAbstract read
In one paragraph

Article in IEEE transactions on ultrasonics, ferroelectrics, and frequency control, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 11 papers.

0numbers the graph read from it
0cells of the map it votes in
11citing papers in PubMed
1.5field-weighted citation impact, top 18% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

11 citing papers in PubMed, 19 citations in OpenAlex.

  1. Review
  2. Review
  3. Article
  4. Article
  5. Article
  6. Review
  7. Article
  8. Blood Flow Imaging in the Neonatal Brain Using Angular Coherence Power Doppler.IEEE transactions on ultrasonics, ferroelectrics, and frequency control · 2021
    Article
  9. A Spatial Coherence Beamformer Design for Power Doppler Imaging.IEEE transactions on medical imaging · 2020
    Article
  10. Article
  11. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 2 institutions in 1 country.

Jaime Tierney
Kristy Walsh
Helen Griffith
Jennifer Baker
Daniel B Brown
Brett Byram
Vanderbilt University · USVanderbilt University Medical Center · US

Funding

Mechanobiology of Cardiopulmonary FibrosisR35HL135790 · NHLBI · VANDERBILT UNIVERSITY · PI MERRYMAN, WILLIAM D · 2017 to 2023
$5.5M
Institutional Predoctoral Training Program in Biomedical Imaging ScienceT32EB014841 · NIBIB · VANDERBILT UNIVERSITY · PI GORE, JOHN C · 2013 to 2017
$771k
Parallel Computer with High Memory NodesS10OD016216 · OD · VANDERBILT UNIVERSITY · PI PISTON, DAVID W · 2013 to 2013
$598k
NHLBI NIH HHS R35 HL135790NIBIB NIH HHS T32 EB014841NIH HHS S10 OD016216
6 · The paper itself

Abstract

Noncontrast perfusion ultrasound imaging remains challenging due to spectral broadening of the tissue clutter signal caused by patient and sonographer hand motion. To address this problem, we previously introduced an adaptive demodulation scheme to suppress the bandwidth of tissue prior to high-pass filtering. Our initial implementation used single plane wave power Doppler imaging and a conventional tissue filter. Recent advancements in beamforming and tissue filtering have been proposed for improved slow flow imaging, including coherent flow power Doppler (CFPD) imaging and singular value decomposition (SVD) filtering. Here, we aim to evaluate adaptive demodulation in conjunction with improvements in beamforming and filtering using simulations, single-vessel phantoms, and an in vivo liver tumor embolization study. We show that simulated blood-to-background contrast-to-noise ratios are highest when using adaptive demodulation with CFPD and a 100-ms ensemble, which resulted in a 13.6-dB average increase in contrast-to-noise ratio compared to basic IIR filtering alone. We also show that combining adaptive demodulation with SVD and with CFPD + SVD results in 9.3- and 19-dB increases in contrast-to-noise ratios compared to IIR filtering alone at 700- and 500-ms ensembles for phantom data with 1- and 5-mm/s average flows, respectively. In general, combining techniques resulted in higher signal-to-noise, contrast-to-noise, and generalized contrast-to-noise ratios in both simulations and phantoms. Finally, adaptive demodulation with SVD resulted in the largest qualitative and quantitative changes in tumor-to-background contrast postembolization.

Indexed as

Signal Processing, Computer-AssistedHumansImage Processing, Computer-AssistedLiverLiver NeoplasmsPerfusion ImagingPhantoms, ImagingUltrasonography

Identifiers

PMID30735994
PMCPMC6528792
OpenAlexW2913653812

What OpenQuestion holds

Textmetadata
LicenceTDM
Read underepoch 390

Registered trials

None linked