Evidence map›Paper›PMID 30712432›Full record

ArticleJournal of the American Heart Association2019

How Well Do Results From Randomized Clinical Trials and/or Recommendations for Implantable Cardioverter-Defibrillator Programming Diffuse Into Clinical Practice?

Niraj Varma, Paul Jones, Nicholas Wold, Edmond Cronin, Kenneth Stein

Open access · goldAbstract readMulticenter Study
In one paragraph

Article in Journal of the American Heart Association, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.

0numbers the graph read from it
0cells of the map it votes in
7citing papers in PubMed
1.9field-weighted citation impact, top 13% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

7 citing papers in PubMed, 17 citations in OpenAlex.

  1. Incidence and risk factors for first and recurrent ICD shock therapy in patients with an implantable cardioverter defibrillator.Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2025
    Article
  2. Article
  3. Article
  4. Remote monitoring of cardiac implantable electronic devices and disease management.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2023
    Article
  5. Article
  6. Outcomes Following Standardized Implantable Cardioverter-defibrillator Reprogramming.The Journal of innovations in cardiac rhythm management · 2022
    Article
  7. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 3 institutions in 1 country.

Niraj Varma1 Cleveland Clinic Cleveland OH.
Paul Jones2 Boston Scientific St. Paul MN.
Nicholas Wold2 Boston Scientific St. Paul MN.
Edmond Cronin3 Hartford Hospital Hartford CT.
Kenneth Stein2 Boston Scientific St. Paul MN.
Boston Scientific (United States) · USCleveland Clinic · USHartford Hospital · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background Inappropriate implantable cardioverter-defibrillator programming can be detrimental. Whether trials/recommendations informing best implantable cardioverter-defibrillator programming (high-rate cutoff and/or extended duration of detection) influence practice is unknown. Methods and Results We measured reaction to publication of MADIT-RIT (Multicenter Automatic Defibrillator Implantation Trial-Reduce Inappropriate Therapy; 2012) and the Consensus Statement (2015) providing generic programming parameters, in a national cohort of implantable cardioverter-defibrillator recipients, using the ALTITUDE database (Boston Scientific). Yearly changes in programmed parameters to either trial-specified or class 1 recommended parameters (≥185 beats per minute or delay ≥6 seconds) were assessed in parallel. From 2008 to 2017, 232 982 patients (aged 67±13 years; 28% women) were analyzed. Prevalence of MADIT- RIT -specific settings before publication was <1%, increasing to 13.6% in the year following. Thereafter, this increased by <6% over 5 years. Among preexisting implants (91 171), most patients (58 739 [64.4%]) underwent at least 1 in-person device reprogramming after trial publication, but <2% were reprogrammed to MADIT - RIT settings. Notably, prevalence of programming to ≥185 beats per minute or delay ≥6 seconds was increased by MADIT - RIT (57.4% in 2013 versus 40.2% at baseline), but the following publication of recommendations had minor incremental effect (73.2% in 2016 versus 70.8% in 2015). High-rate cutoff programming was favored almost 2-fold compared with extended duration throughout the test period. Practice changes demonstrated large interhospital and interstate variations. Conclusions Trial publication had an immediate effect during 1 year postpublication, but absolute penetration was low, and amplified little with time. Consensus recommendations had a negligible effect. However, generic programming was exercised more widely, and increased after trial publication, but not following recommendations.

Indexed as

Defibrillators, ImplantablePractice Guidelines as TopicAgedEquipment FailureFollow-Up StudiesHumansRandomized Controlled Trials as TopicRetrospective StudiesRisk FactorsTachycardia, VentricularTime Factorsguideline adherenceguidelinesimplantable cardioverter‐defibrillatorprogrammingremote monitoring

Identifiers

PMID30712432
PMCPMC6405582
OpenAlexW2914053432

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked