ArticleCureus2018

Evaluation of CyberKnife® Fiducial Tracking Limitations to Assist Targeting Accuracy: A Phantom Study with Fiducial Displacement.

Christy Goldsmith, Melanie M Green, Brownwyn Middleton, Ian Cowley, Andrew Robinson, Nicholas P Plowman, Patricia M Price

Abstract read
In one paragraph

Article in Cureus, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 11 papers.

0numbers the graph read from it
0cells of the map it votes in
11citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

11 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
  9. Article
  10. Review
  11. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Christy GoldsmithRadiation Oncology, The Cyberknife Centre, HCA Healthcare, London, GBR.
Melanie M GreenRadiation Oncology, Imperial College London, London, GBR.
Brownwyn MiddletonRadiation Oncology, The Harley Street Clinic, HCA Healthcare, London, GBR.
Ian CowleyMedical Physics, The Harley Street Clinic, HCA Healthcare, London, GBR.
Andrew RobinsonMedical Physics, The Harley Street Clinic, HCA Healthcare, London, GBR.
Nicholas P PlowmanRadiation Oncology, The Harley Street Clinic, HCA Healthcare, London, GBR.
Patricia M PriceRadiation Oncology, The Harley Street Clinic, HCA Healthcare, London, GBR.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Introduction  The underlying assumptions of the CyberKnife® (Accuray, Sunnyvale, CA, US) fiducial tracking system are: i) fiducial positions are accurately detected; ii) inter-fiducial geometry remains consistent (rigid); iii) inter-fiducial geometric array changes are detected and either accommodated with corrections or treatment is interrupted. However: i) soft-tissue targets are deformable & fiducial migration is possible; ii) the accuracy of the tracking system has not previously been examined with fiducial displacement; iii) treatment interruptions may occur due to inter-fiducial geometric changes, but there is little information available to assist subsequent troubleshooting. The purpose of this study was to emulate a clinical target defined with a two, three, or four-fiducial array where one fiducial is displaced to mimic a target deformation or fiducial migration scenario. The objectives: evaluate the fiducial positioning accuracy, array interpretation, & corresponding corrections of the CyberKnife system, with the aim of assisting troubleshooting following fiducial displacement. Methods A novel solid-water phantom was constructed with three fixed fiducials (F1,F2,F3) & one moveable fiducial (F4), arranged as if placed to track an imaginary clinical target. Using either two fiducials (F1,F4), different combinations of three fiducials (F1,F2,F4; F1,F3,F4; F2,F3,F4) or four fiducials (F1,F2,F3,F4), repeat experiments were conducted where F4 was displaced inferiorly at 2-mm intervals from 0-16 mm. Data were acquired at each position of F4, including rigid body errors (RBE), fiducial x, y, & z coordinate displacements, six degrees of freedom (DOF) corrections, & robot center-of-mass (COM) translation corrections. Results Maximum positioning difference (mean±SD) between the reference and live x, y, & z coordinates for the three fixed fiducials was 0.08±0.30 mm, confirming good accuracy for fixed fiducial registration. For two fiducials (F1,F4), F4 registration was accurate to 14-mm displacement and the F4 x-axis coordinate change was 2.0±0.12 mm with each 2 mm inferior displacement validating the phantom for tracking evaluation. RBE was >5 mm (system threshold) at 6-14 mm F4 displacement: however, F1 was misidentified as the RBE main contributor. Further, F1/F4 false-lock occurred at 16 mm F4 displacement with corresponding RBE <3 mm & COM corrections >13 mm. For combinations of three fiducials, F4 registration was accurate to 10-mm displacement. RBE was >5 mm at 6-16 mm F4 displacement: however, F4 false-lock occurred at 12-16 mm with RBE 5-6 mm. For four fiducials, F4 registration was accurate to 4 mm displacement: however, F4 false-lock occurred at 6-16 mm displacement with concerning RBE <2 & <5 at 6 & 8-mm F4 displacement, respectively. False-locks were easily identified in the phantom but frequently uncorrectable. Conclusions Results indicate fiducial positioning accuracy and system output following fiducial displacement depends on the number of fiducials correlated, displacement distance, and clinical thresholds applied. Displacements ≤4 mm were accurately located, but some displacements 6-16 mm were misrepresented, either by erroneous main contributor (two-fiducial array only) or by false-locks and misleading RBE, which underestimated displacement. Operator vigilance and implementation of our practical guidelines based on the study findings may help reduce targeting error and assist troubleshooting in clinical situations.

Indexed as

accuracycyberknifefiducial markerslimitationsphantomtarget motiontroubleshootingtumor tracking

Identifiers

PMID30648058
PMCPMC6318119

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.