Evidence map›Paper›PMID 30633548›Full record

ArticleThe British journal of radiology2019

The associated factors for radiation dose variation in cardiac CT angiography.

Ali B Alhailiy, Ernest U Ekpo, Peter L Kench, Elaine A Ryan, Patrick C Brennan, Mark McEntee

Abstract readMulticenter Study
In one paragraph

Article in The British journal of radiology, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed
0.9field-weighted citation impact, top 26% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed, 10 citations in OpenAlex.

  1. Article
  2. Trading off Iodine and Radiation Dose in Coronary Computed Tomography.Journal of cardiovascular development and disease · 2025
    Review
  3. Article
  4. Article
  5. Article
  6. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 1 institution in 2 countries.

Ali B Alhailiy1 The University of Sydney, Faculty of Health Sciences, Discipline of Medical Radiation Science , Sydney, NSW , Australia.ORCID http://orcid.org/0000-0001-5683-7115
Ernest U Ekpo1 The University of Sydney, Faculty of Health Sciences, Discipline of Medical Radiation Science , Sydney, NSW , Australia.
Peter L Kench1 The University of Sydney, Faculty of Health Sciences, Discipline of Medical Radiation Science , Sydney, NSW , Australia.
Elaine A Ryan1 The University of Sydney, Faculty of Health Sciences, Discipline of Medical Radiation Science , Sydney, NSW , Australia.
Patrick C Brennan1 The University of Sydney, Faculty of Health Sciences, Discipline of Medical Radiation Science , Sydney, NSW , Australia.
Mark McEntee1 The University of Sydney, Faculty of Health Sciences, Discipline of Medical Radiation Science , Sydney, NSW , Australia.
The University of Sydney · AU

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objective: This study aimed to examine the associated factors for dose variation and influence cardiac CT angiography (CCTA) dose benchmarks in current CT imaging centres.

methods: A questionnaire was distributed to CT centres across Australia and Saudi Arabia. All participating centres collected data for adults who underwent a CCTA procedure. The questionnaire gathered information about the examination protocol, scanning parameters, patient parameters, and volume CT dose index (CTDI vol) and dose-length product (DLP). A stepwise regression analysis was performed to assess the contribution of tube voltage (kV), padding time technique, cross-sectional area (CSA) of chest and weight to DLP.

results: A total of 17 CT centres provided data for 423 CCTA examinations. The median CTDI

conclusion: Findings show radiation dose variations for CCTA. The associated factors for dose variation found in this study are scanning mode, kV, padding time technique and CSA of the chest. This results support the need to include CSA measurements in future dose survey and for setting DRLs. ADVANCES IN KNOWLEDGE:: The study provides baseline information that helps to understand the associated factors for dose variations and high doses within and between centres performing CCTA.

Indexed as

Radiation DosageAustraliaComputed Tomography AngiographyCoronary AngiographyFemaleHumansMaleMiddle AgedSaudi ArabiaSurveys and Questionnaires

Identifiers

PMID30633548
PMCPMC6886289
OpenAlexW2909284349

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked