Evidence map›Paper›PMID 30604307›Full record

ArticleJournal of cardiovascular translational research2019

Evaluation of the CorInnova Heart Assist Device in an Acute Heart Failure Model.

Erica C Hord, Christina M Bolch, Egemen Tuzun, William E Cohn, Boris Leschinsky, John C Criscione

Abstract read
In one paragraph

Article in Journal of cardiovascular translational research, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 12 papers.

0numbers the graph read from it
0cells of the map it votes in
12citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

12 citing papers in PubMed.

  1. Review
  2. Article
  3. Article
  4. Review
  5. Review
  6. Article
  7. Review
  8. Article
  9. Review
  10. Article
  11. Article
  12. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Erica C HordDepartment of Biomedical Engineering, Texas A&M University, Mailstop 3120, College Station, TX, 77843-3120, USA.
Christina M BolchCorInnova, Incorporated, Houston, TX, USA.
Egemen TuzunTexas A&M Institute for Preclinical Studies, Texas A&M University, College Station, TX, USA.
William E CohnTexas Heart Institute, Houston, TX, USA.
Boris LeschinskyCorInnova, Incorporated, Houston, TX, USA.
John C CriscioneDepartment of Biomedical Engineering, Texas A&M University, Mailstop 3120, College Station, TX, 77843-3120, USA. jccriscione@tamu.edu.ORCID 0000-0002-5988-773X

Funding

Wellcome Trust 104613
6 · The paper itself

Abstract

While the number of patients supported with temporary cardiac assist is growing, the existing devices are limited by a multitude of complications, mostly related to contact with the blood. The CorInnova epicardial compressive heart assist device was tested in six sheep using an acute heart failure model. High esmolol dose, targeting a 50% reduction in CO from healthy baseline, resulted in a failure state with mean CO 1.9 L/min. Heart assist with the device during failure state resulted in an average absolute increase in CO of 1.0 L/min, along with a decline in ventricular work to 67.5% of the total LV SW. Combined with repeated success of minimally invasive device implant, the resulting increases in cardiac hemodynamics achieved while still unloading the heart demonstrate the potential of the CorInnova device for temporary heart assist.

Indexed as

Heart-Assist DevicesHemodynamicsVentricular Function, LeftAnimalsDisease Models, AnimalHeart FailureMaterials TestingProsthesis DesignProsthesis ImplantationRecovery of FunctionSheep, DomesticAcute failureCardiac devicesCardiac unloadingCorInnovaDirect cardiac compressionHeart assistTemporary support

Identifiers

PMID30604307
PMCPMC6497617

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked