A compact pulsatile simulator based on cam-follower mechanism for generating radial pulse waveforms.
Article in Biomedical engineering online, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
8 citing papers in PubMed, 52 citations in OpenAlex.
- Development of a high-precision evaluation system for radial pulse wave applanation tonometry devices.Scientific reports · 2026Article
- Continuous dynamic microforce reconstruction using electrical stimulation for remote pulse diagnosis.Fundamental research · 2025Article
- An Open Source Validation System for Continuous Arterial Blood Pressure Measuring Sensors.Sensors (Basel, Switzerland) · 2025Article
- Research and development of the pulse acquisition system and the pulse biomimetic reproduction system.Journal of traditional Chinese medicine = Chung i tsa chih ying wen pan · 2025Article
- CT-Derived 3D Printing for Coronary Artery Cannulation Simulator Design Manufacturing.Bioengineering (Basel, Switzerland) · 2022Article
- A Transfer Function Model Development for Reconstructing Radial Pulse Pressure Waveforms Using Non-Invasively Measured Pulses by a Robotic Tonometry System.Sensors (Basel, Switzerland) · 2021Article
- Portable Interactive Pulse Tactile Recorder and Player System.Sensors (Basel, Switzerland) · 2021Article
- Measuring Arterial Pulsatility With Dynamic Inflow Magnitude Contrast.Frontiers in neuroscience · 2021Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors at 5 institutions in 2 countries.
Funding
Abstract
backgroundThere exists a growing need for a cost-effective, reliable, and portable pulsation simulator that can generate a wide variety of pulses depending on age and cardiovascular disease. For constructing compact pulsation simulator, this study proposes to use a pneumatic actuator based on cam-follower mechanism controlled by a DC motor. The simulator is intended to generate pulse waveforms for a range of pulse pressures and heart beats that are realistic to human blood pulsations.
methodsThis study first performed in vivo testing of a healthy young man to collect his pulse waveforms using a robotic tonometry system (RTS). Based on the collected data a representative human radial pulse waveform is obtained by conducting a mathematical analysis. This standard pulse waveform is then used to design the cam profile. Upon fabrication of the cam, the pulsatile simulator, consisting of the pulse pressure generating component, pressure and heart rate adjusting units, and the real-time pulse display, is constructed. Using the RTS, a series of testing was performed on the prototype to collect its pulse waveforms by varying the pressure levels and heart rates. Followed by the testing, the pulse waveforms generated by the prototype are compared with the representative, in vivo, pulse waveform.
resultsThe radial Augmentation Index analysis results show that the percent error between the simulator data and human pulse profiles is sufficiently small, indicating that the first two peak pressures agree well. Moreover, the phase analysis results show that the phase delay errors between the pulse waveforms of the prototype and the representative waveform are adequately small, confirming that the prototype simulator is capable of simulating realistic human pulse waveforms.
conclusionsThis study demonstrated that a very accurate radial pressure waveform can be reproduced using the cam-based simulator. It can be concluded that the same testing and design methods can be used to generate pulse waveforms for other age groups or any target pulse waveforms. Such a simulator can make a contribution to the research efforts, such as development of wearable pressure sensors, standardization of pulse diagnosis in oriental medicine, and training medical professionals for pulse diagnosis techniques.