Evidence map›Paper›PMID 30595271›Full record

Trial reportIndian heart journal2018

Efficacy of RADPAD protection drape in reducing radiation exposure in the catheterization laboratory-First Indian study.

Pujan Shah, Roopali Khanna, Aditya Kapoor, Pravin K Goel

Open access · goldAbstract readRandomized Controlled Trial
In one paragraph

Trial report in Indian heart journal, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed
1.1field-weighted citation impact, top 22% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed, 21 citations in OpenAlex.

  1. Trial
  2. Article
  3. Article
  4. Review
  5. Article
  6. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors at 1 institution in 1 country.

Pujan ShahSanjay Gandhi PGIMS, Lucknow, India.
Roopali KhannaSanjay Gandhi PGIMS, Lucknow, India.
Aditya KapoorSanjay Gandhi PGIMS, Lucknow, India.
Pravin K GoelSanjay Gandhi PGIMS, Lucknow, India. Electronic address: golf_pgi@yahoo.co.in.
Sanjay Gandhi Post Graduate Institute of Medical Sciences · IN

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundOccupational radiation exposureis a growing problem due to increasing number and complexity of interventional procedures.The RADPAD is a lead-free sterile drape containing bismuth and barium that reduces scatter radiation during fluoroscopic procedures. We aimed to study the radiation exposure reduction to operators with the use of RADPAD and also measureradiation doses in different angiographic projections.

methods65 randomly selected patients undergoing elective complex percutaneous coronary intervention (PCI) procedures from January 2017 to 2017 were randomized in a 1:1 pattern with or without the RADPAD. Primary endpoint was the ratio of operator received dose in mrem to total radiation in Gyat the end of the procedure which was designated ''Relative operator exposure'', with or without RADPAD.

resultsDespite similar fluoroscopy times (20.4 ± 9.4 min with RADPAD vs. 19.4 ± 9.2 min without RADPAD, P = 0.871) and total radiation dose (3.4 ± 4.3 Gy with RADPAD vs. 2.3 ± 1.4 Gy, P = 0.198), the relative operator exposure was significantly less with RADPAD (1.39 ± 0.95) as compared to no RADPAD group (2.27 ± 1.4) (p = 0.004) amounting to a 39% reduction. Additionally mean radiation dose per shoot of recorded Left anterior oblique (LAO) oriented projections was 34.4 ± 15.7mGyvs 24.9 ± 12.9 mGy for a non LAO oriented projection. (p < 0.001).

conclusionRADPAD significantly reduces radiation exposure to the primary operator during prolonged complex PCI procedures. Further, amongst all views, LAO views have significantly higher emitted radiation as compared to Non LAO views and need more radiation protection.

Indexed as

Protective DevicesCardiac CatheterizationCoronary AngiographyCoronary Artery DiseaseEquipment DesignFemaleFluoroscopyFollow-Up StudiesHumansIncidenceIndiaMaleMiddle AgedPercutaneous Coronary InterventionRadiation InjuriesRadiation ProtectionChronic total occlusionLeft anterior obliqueLeft main arteryPercutaneous coronary interventionRoentgen equivalent man