Evidence map›Paper›PMID 30588134›Full record

ArticleMedical devices (Auckland, N.Z.)2018

Preclinical quantification of air leaks in a physiologic lung model: effects of ventilation modality and staple design.

Chad E Eckert, Jason L Harris, Jordan B Wong, Suzanne Thompson, Edmund S Kassis, Masahiro Tsuboi, Harald C Ott, Seth Force

Abstract read
In one paragraph

Article in Medical devices (Auckland, N.Z.), 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed.

  1. Article
  2. Ex-PALM: a negative pressure ventilation system for air leak evaluation inFrontiers in bioengineering and biotechnology · 2025
    Article
  3. Review
  4. Sound-guided assessment and localization of pulmonary air leak.Bioengineering & translational medicine · 2023
    Article
  5. Article
  6. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors.

Chad E EckertEthicon Inc., Research and Development, Cincinnati, OH 45242, USA, ceckert1@its.jnj.com.
Jason L HarrisEthicon Inc., Research and Development, Cincinnati, OH 45242, USA, ceckert1@its.jnj.com.
Jordan B WongEthicon Inc., Research and Development, Cincinnati, OH 45242, USA, ceckert1@its.jnj.com.
Suzanne ThompsonEthicon Inc., Preclinical Center of Excellence, Cincinnati, OH 45242, USA.
Edmund S KassisEthicon Inc., Medical Affairs, Cincinnati, OH 45242, USA.
Masahiro TsuboiDivision of Thoracic Surgery and Oncology, National Cancer Center Hospital East, Kashiwanoha, Kashiwa, Chiba 277-8577, Japan.
Harald C OttDepartment of Surgery, Division of Thoracic Surgery, Massachusetts General Hospital, Boston, MA 02114, USA.
Seth ForceDepartment of Surgery, Division of Cardiothoracic Surgery, Emory University Hospital, Atlanta, GA 30322, USA.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeThoracic air leaks are a common complication following pulmonary resections. Limitations in clinical studies and preclinical models have hindered efforts to understand the pathophysiology of air leaks. With an emphasis on staple-line specific air leaks, we hypothesize that ventilation modality - intraoperative positive pressure vs postoperative negative pressure - and stapler design may play a role in air leaks.

methodsUsing a novel physiologic lung model, air leaks associated with graduated and uniform staple designs were evaluated under positive and negative pressure ventilation, simulating perioperative breathing in porcine lungs. Air leak incidence, air leak volume, and air leak rate were captured along with ventilation pressure and tidal volume.

resultsIn all cases, negative pressure ventilation was associated with a higher occurrence of leaks when compared to positive pressure ventilation. Lungs leaked more air and at a faster rate under negative pressure ventilation compared to positive pressure ventilation. Graduated staple designs were associated with higher occurrence of leaks as well as larger leak rates when compared to uniform staples. Tissue thickness was not associated with differences in air leaks when tested with appropriate staple heights.

conclusionUsing a novel lung model to investigate the pathophysiology of air leaks, we have identified breathing modality and staple design as two important variables that may impact air leaks. This work will help guide device design and drive future studies in human tissue, and it may help inform clinical practice to ultimately improve patient outcomes.

Indexed as

lung physiologypreclinical modelprolonged air leaksurgical staplerventilation mechanics

Identifiers

PMID30588134
PMCPMC6300381

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.