ArticleJournal of applied clinical medical physics2019
Automatic measurement of air gap for proton therapy using orthogonal x-ray imaging with radiopaque wires.
Article in Journal of applied clinical medical physics, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed, 1 citations in OpenAlex.
- A novel approach to Verify air gap and SSD for proton radiotherapy using surface imaging.Radiation oncology (London, England) · 2019Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
10 authors at 5 institutions in 2 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeThe main objective of this study was to develop a technique to accurately determine the air gap between the end of the proton beam compensator and the body of the patient in proton radiotherapy.
methodsOrthogonal x-ray image-based automatic coordinate reconstruction was used to determine the air gap between the patient body surface contour and the end of beam nozzle in proton radiotherapy. To be able to clearly identify the patient body surface contour on the orthogonal images, a radiopaque wire was placed on the skin surface of the patient as a surrogate. In order to validate this method, a Rando
resultsOn the phantom study, the air gap values derived with the automatic reconstruction method were found to be within 1.1 mm difference from the planned values for proton beams with air gaps of 85.0, 100.0, 150.0, 180.0, and 200.0 mm. The reconstruction technique determined air gaps for a patient in two clinical treatment sessions were 38.4 and 41.8 mm, respectively, for a 40 mm planned air gap, and confirmed by manual measurements. There was strong agreement between the calculated values and the automatically measured values, and between the automatically and manually measured values.
conclusionsAn image-based automatic method has been developed to conveniently determine the air gap of a proton beam, directly using the orthogonal images for patient positioning without adding additional imaging dose to the patient. The method provides an objective, accurate, and efficient way to confirm the target depth at treatment to ensure desired target coverage and normal tissue sparing.