Evidence map›Paper›PMID 30518105›Full record

ArticleBiosensors2018

Evaluation of Continuous Lactate Monitoring Systems within a Heparinized In Vivo Porcine Model Intravenously and Subcutaneously.

Alexander Wolf, Kevin Renehan, Kenneth K Y Ho, Benjamin D Carr, Chieh V Chen, Marie S Cornell, Minyi Ye, Alvaro Rojas-Peña, Hao Chen

Open access · goldAbstract readComparative Study
In one paragraph

Article in Biosensors, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.

0numbers the graph read from it
0cells of the map it votes in
5citing papers in PubMed
1.0field-weighted citation impact, top 26% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

5 citing papers in PubMed, 19 citations in OpenAlex.

  1. Serial Lactate in Clinical Medicine - A Narrative Review.Journal of intensive care medicine · 2026
    Review
  2. Real-Time,ACS sensors · 2022
    Review
  3. Article
  4. Review
  5. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors at 2 institutions in 1 country.

Alexander WolfBiocrede Inc., Plymouth, MI 48170, USA. alexander.wolf@biocrede.com.
Kevin RenehanBiocrede Inc., Plymouth, MI 48170, USA. kevin.renehan@biocrede.com.
Kenneth K Y HoBiocrede Inc., Plymouth, MI 48170, USA. kenneth.ho@biocrede.com.
Benjamin D CarrDepartment of Surgery-ECLS Laboratory, University of Michigan Medical School, Ann Arbor, MI 48109, USA. bdecarr@umich.edu.ORCID 0000-0002-3563-8974
Chieh V ChenBiocrede Inc., Plymouth, MI 48170, USA. chieh.07@gmail.com.
Marie S CornellDepartment of Surgery-ECLS Laboratory, University of Michigan Medical School, Ann Arbor, MI 48109, USA. mhaeussl@med.umich.edu.
Minyi YeBiocrede Inc., Plymouth, MI 48170, USA. miya.ye@biocrede.com.
Alvaro Rojas-PeñaDepartment of Surgery-ECLS Laboratory, University of Michigan Medical School, Ann Arbor, MI 48109, USA. alvaror@umich.edu.ORCID 0000-0002-6197-8892
Hao ChenBiocrede Inc., Plymouth, MI 48170, USA. hao.chen@biocrede.com.
Biocrede (United States) · USUniversity of Michigan · US

Funding

Bactericidal Intravascular Biosensor for Continuous Reading of Lactate in the lCU Ph2R44GM119883 · NIGMS · BIOCREDE, INC. · PI CHEN, HAO · 2018 to 2019
$1.4M
Bactericidal Intravascular Biosensor for Continuous Reading of Lactate in the ICUR43GM119883 · NIGMS · BIOCREDE, INC. · PI CHEN, HAO · 2016 to 2016
$202k
NIGMS NIH HHS R43 GM119883NIGMS NIH HHS R44 GM119883NIH HHS 1R43GM119883-01
6 · The paper itself

Abstract

We present an animal model used to evaluate the in vivo performance of electrochemical amperometric continuous lactate sensors compared to blood gas instruments. Electrochemical lactate sensors were fabricated, placed into 5 Fr central venous catheters (CVCs), and paired with wireless potentiostat devices. Following in vivo evaluation and calibration, sensors were placed within the jugular and femoral veins of a porcine subject as a preliminary assessment of in vivo measurement accuracy. The mobile electronic circuit potentiostat devices supplied the operational voltage for the sensors, measured the resultant steady-state current, and recorded the sensor response values in internal memory storages. An in vivo time trace of implanted intravenous (IV) sensors demonstrated lactate values that correlated well with the discrete measurements of blood samples on a benchtop point-of-care sensor-based instrument. Currents measured continuously from the implanted lactate sensors over 10 h were converted into lactate concentration values through use of a two-point in vivo calibration. Study shows that intravenously implanted sensors had more accurate readings, faster peak-reaching rates, and shorter peak-detection times compared to subcutaneously placed sensors. IV implanted and subcutaneously placed sensors closer to the upper body (in this case neck) showed faster response rates and more accurate measurements compared to those implanted in the lower portion of the porcine model. This study represents an important milestone not only towards continuous lactate monitoring for early diagnosis and intervention in neonatal patients with congenital heart disease undergoing cardiopulmonary bypass surgeries, but also in the intervention of critical ill patients in the Intensive Care Units or during complex surgical procedures.

Indexed as

AnimalsBiosensing TechniquesBlood Gas AnalysisCalibrationCentral Venous CathetersFemaleHeparinHumansLactic AcidModels, AnimalMonitoring, PhysiologicSensitivity and SpecificitySwineWireless TechnologyHeparinLactic Acidcardiopulmonary bypass surgeriescongenital heart diseasecontinuous blood lactate monitoringintravenouslactate sensorssubcutaneous

Identifiers

PMID30518105
PMCPMC6316727
OpenAlexW2903091960

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked