The evaluation of a hybrid biomechanical deformable registration method on a multistage physical phantom with reproducible deformation.

An Qin, Dan Ionascu, Jian Liang, Xiao Han, Nicolette O'Connell, Di Yan

Open access · goldAbstract read
In one paragraph

Article in Radiation oncology (London, England), 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed
3.4field-weighted citation impact, top 7% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed, 16 citations in OpenAlex.

  1. Article
  2. Review
  3. Article
  4. Article
  5. Review
  6. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 3 institutions in 1 country.

An QinDepartment of Radiation Oncology, Beaumont Health System, Royal Oak, MI, USA. an.qin@beaumont.edu.ORCID http://orcid.org/0000-0002-3333-5363
Dan IonascuDepartment of Radiation Oncology, College of Medicine, University of Cincinnati, Cincinnati, OH, USA.
Jian LiangDepartment of Radiation Oncology, Beaumont Health System, Royal Oak, MI, USA.
Xiao HanElekta Inc., Maryland Heights, MO, USA.
Nicolette O'ConnellElekta Inc., Maryland Heights, MO, USA.
Di YanDepartment of Radiation Oncology, Beaumont Health System, Royal Oak, MI, USA.
Beaumont Health · USElekta (United States) · USUniversity of Cincinnati Medical Center · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundAdvanced clinical applications, such as dose accumulation and adaptive radiation therapy, require deformable image registration (DIR) algorithms capable of voxel-wise accurate mapping of treatment dose or functional imaging. By utilizing a multistage deformable phantom, the authors investigated scenarios where biomechanical refinement method (BM-DIR) may be better than the pure image intensity based deformable registration (IM-DIR).

methodsThe authors developed a biomechanical-model based DIR refinement method (BM-DIR) to refine the deformable vector field (DVF) from any initial intensity-based DIR (IM-DIR). The BM-DIR method was quantitatively evaluated on a novel phantom capable of ten reproducible gradually-increasing deformation stages using the urethra tube as a surrogate. The internal DIR accuracy was inspected in term of the Dice similarity coefficient (DSC), Hausdorff and mean surface distance as defined in of the urethra structure inside the phantom and compared with that of the initial IM-DIR under various stages of deformation. Voxel-wise deformation vector discrepancy and Jacobian regularity were also inspected to evaluate the output DVFs. In addition to phantom, two pairs of Head&Neck patient MR images with expert-defined landmarks inside parotids were utilized to evaluate the BM-DIR accuracy with target registration error (TRE).

resultsThe DSC and surface distance measures of the inner urethra tube indicated the BM-DIR method can improve the internal DVF accuracy on masked MR images for the phases of a large degree of deformation. The smoother Jacobian distribution from the BM-DIR suggests more physically-plausible internal deformation. For H&N cancer patients, the BM-DIR improved the TRE from 0.339 cm to 0.210 cm for the landmarks inside parotid on the masked MR images.

conclusionsWe have quantitatively demonstrated on a multi-stage physical phantom and limited patient data that the proposed BM-DIR can improve the accuracy inside solid organs with large deformation where distinctive image features are absent.

Indexed as

AlgorithmsPhantoms, ImagingHead and Neck NeoplasmsHumansImage Processing, Computer-AssistedMaleProstateRadiotherapy DosageRadiotherapy, Intensity-ModulatedRadiotherapy Planning, Computer-AssistedTomography, X-Ray ComputedUrethraadaptive radiotherapyBiomechanical modelDeformable image registrationdose accumulationIGRT

Identifiers

PMID30514348
PMCPMC6280462
OpenAlexW2902662464