Evidence map›Paper›PMID 30429115›Full record

ArticleJMIR mHealth and uHealth2018

A Cardiopulmonary Monitoring System for Patient Transport Within Hospitals Using Mobile Internet of Things Technology: Observational Validation Study.

Jang Ho Lee, Yu Rang Park, Solbi Kweon, Seulgi Kim, Wonjun Ji, Chang-Min Choi

Erratum issuedAbstract read
In one paragraph

Article in JMIR mHealth and uHealth, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. An erratum has been issued. Cited by 2 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed, 1 pooled it
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed, 1 synthesis or guideline pooled it.

  1. Pooled it
  2. Review
4 · The record

Corrections and comments

5 · Who and what money

Authors and funding

6 authors.

Jang Ho LeeDepartment of Pulmonology and Critical Care Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Republic of Korea.ORCID http://orcid.org/0000-0002-8226-9585
Yu Rang ParkDepartment of Biomedical System Informatics, Yonsei University College of Medicine, Seoul, Republic of Korea.ORCID http://orcid.org/0000-0002-4210-2094
Solbi KweonDepartment of Biomedical System Informatics, Yonsei University College of Medicine, Seoul, Republic of Korea.ORCID http://orcid.org/0000-0002-8021-9605
Seulgi KimDepartment of Pulmonology and Critical Care Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Republic of Korea.ORCID http://orcid.org/0000-0001-5002-3313
Wonjun JiDepartment of Pulmonology and Critical Care Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Republic of Korea.ORCID http://orcid.org/0000-0001-7164-2770
Chang-Min ChoiDepartment of Pulmonology and Critical Care Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Republic of Korea.ORCID http://orcid.org/0000-0002-2881-4669

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundDuring intrahospital transport, adverse events are inevitable. Real-time monitoring can be helpful for preventing these events during intrahospital transport.

objectiveWe attempted to determine the viability of risk signal detection using wearable devices and mobile apps during intrahospital transport. An alarm was sent to clinicians in the event of oxygen saturation below 90%, heart rate above 140 or below 60 beats per minute (bpm), and network errors. We validated the reliability of the risk signal transmitted over the network.

methodsWe used two wearable devices to monitor oxygen saturation and heart rate for 23 patients during intrahospital transport for diagnostic workup or rehabilitation. To determine the agreement between the devices, records collected every 4 seconds were matched and imputation was performed if no records were collected at the same time by both devices. We used intraclass correlation coefficients (ICC) to evaluate the relationships between the two devices.

resultsData for 21 patients were delivered to the cloud over LTE, and data for two patients were delivered over Wi-Fi. Monitoring devices were used for 20 patients during intrahospital transport for diagnostic work up and for three patients during rehabilitation. Three patients using supplemental oxygen before the study were included. In our study, the ICC for the heart rate between the two devices was 0.940 (95% CI 0.939-0.942) and that of oxygen saturation was 0.719 (95% CI 0.711-0.727). Systemic error analyzed with Bland-Altman analysis was 0.428 for heart rate and -1.404 for oxygen saturation. During the study, 14 patients had 20 risk signals: nine signals for eight patients with less than 90% oxygen saturation, four for four patients with a heart rate of 60 bpm or less, and seven for five patients due to network error.

conclusionsWe developed a system that notifies the health care provider of the risk level of a patient during transportation using a wearable device and a mobile app. Although there were some problems such as missing values and network errors, this paper is meaningful in that the previously mentioned risk detection system was validated with actual patients.

Indexed as

heart rateintrahospital transportmobile applicationoxygen saturationpatient safetyreal-time monitoringwearable device

Identifiers

PMID30429115
PMCPMC6262206

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.