ArticlePractical laboratory medicine2018
Plasma lactate measurement as an example of encountered gaps between routine clinical laboratory processes and manufactures' sample-handling instructions.
Article in Practical laboratory medicine, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- Article
- Hypophosphatemia in suspected seizures evaluated in first seizure clinics and neurology consults.Epilepsia open · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectivesDeviation from manufacturers' pre-analytical sample handling recommendations necessitates extensive validation studies. This report uses plasma lactate testing, where a recommended 15 min room temperature sample handling limit cannot be met by the clinical laboratory, as an example for studies to bridge the gap with practice. DESIGN AND
methodsTriplicate blood samples were collected from patients (n = 51) with lactate requests by clinicians and from normal volunteers (n = 50). One tube was transported on ice (4 °C), the others were maintained at room temperature (23 °C). Tubes stored at 4 °C were processed at 30 min from collection. Tubes stored at 23 °C were processed at 15 and at 30 min from collection. Lactate levels were measured using Roche Diagnostics Cobas 6000® analyzer.
resultsLactate levels in normal subjects ranged from 0.6 to 3.1 mmol/L (median 1.1). Patient lactate levels ranged from 0.8 to 26.3 mmol/L (median 2.2). Bias in lactate levels following extended storage of samples from both normal subjects and patients ranged from - 1.3 to2.2 and from - 1.0-1.0 mmol/L when stored for 30 min at 23 °C or at 4 °C, respectively. The bias between lactate levels at 30 min at 23 °C and 4 °C was - 1.2 to - 0.5 mmol/L for both populations. Although the bias was not statistically significant for all variables, a clinically significant (>0.2 mmol/L) bias was observed in 28% of normal and 7.0% of patient samples.
conclusionExtending the pre-analytical time to 30 min at 23 °C did not significantly impact clinical utility of lactate measurement in our patient population.