Signal and noise characteristics of a CdTe-based photon counting detector: Cascaded systems analysis and experimental studies.

Xu Ji, Ran Zhang, Yongshuai Ge, Guang-Hong Chen, Ke Li

Abstract read
In one paragraph

Article in Proceedings of SPIE--the International Society for Optical Engineering, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.

0numbers the graph read from it
0cells of the map it votes in
5citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

5 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Analogous Lubberts effect in photon counting detectors.Proceedings of SPIE--the International Society for Optical Engineering · 2020
    Article
  5. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Xu JiDepartment of Medical Physics, University of Wisconsin School of Medicine and Public Health, Madison, WI.
Ran ZhangDepartment of Medical Physics, University of Wisconsin School of Medicine and Public Health, Madison, WI.
Yongshuai GeDepartment of Medical Physics, University of Wisconsin School of Medicine and Public Health, Madison, WI.
Guang-Hong ChenDepartment of Medical Physics, University of Wisconsin School of Medicine and Public Health, Madison, WI.
Ke LiDepartment of Medical Physics, University of Wisconsin School of Medicine and Public Health, Madison, WI.

Funding

Multi-Contrast X-ray Breast ImagingR01EB020521 · NIBIB · UNIVERSITY OF WISCONSIN-MADISON · PI CHEN, GUANG-HONG, LI, KE · 2016 to 2020
$2.0M
NIBIB NIH HHS R01 EB020521
6 · The paper itself

Abstract

Recent advances in single photon counting detectors (PCDs) are opening up new opportunities in medical imaging. However, the performance of PCDs is not flawless. Problems such as charge sharing may deteriorate the performance of PCD. This work studied the dependence of the signal and noise properties of a cadmium telluride (CdTe)-based PCD on the charge sharing effect and the anti-charge sharing (ACS) capability offered by the PCD. Through both serial and parallel cascaded systems analysis, a theoretical model was developed to trace the origin of charge sharing in CdTe-based PCD, which is primarily related to remote k-fluorescence re-absorption and spatial spreading of charge cloud. The ACS process was modeled as a sub-imaging state prior to the energy thresholding stage, and its impact on the noise power spectrum (NPS) of PCD can be qualitatively determined by the theoretical model. To validate the theoretical model, experimental studies with a CdTe-based PCD system (XC-FLITE X1, XCounter AB) was performed. Two x-ray radiation conditions, including an RQA-5 beam and a 40 kVp beam, were used for the NPS measurements. Both theoretical predictions and experimental results showed that ACS makes the NPS of the CdTe-based PCD flatter, which corresponds to reduced noise correlation length. The flatness of the NPS is further boosted by increasing the energy threshold or reducing the x-ray energy, both of which reduce the likelihood of registering multiple counts from the same incidenting x-ray photon.

Indexed as

anti charge sharingcascaded systems analysisCdTecharge sharingnoise power spectrumphoton counting detectorX-ray detectorx-ray imaging

Identifiers

PMID30416244
PMCPMC6225770

What OpenQuestion holds

Textmetadata
LicenceTDM
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.