Evidence map›Paper›PMID 30350020›Full record

ArticleEuropean journal of orthopaedic surgery & traumatology : orthopedie traumatologie2019

Intraoperative kinematic analysis of posterior stabilized total knee arthroplasty with asymmetric helical post-cam design.

Takatoshi Morooka, Makiko Okuno, Daisuke Seino, Takuya Iseki, Shigeo Fukunishi, Syoji Kobashi, Shinichi Yoshiya

Abstract read
PubMed Publisher
In one paragraph

Article in European journal of orthopaedic surgery & traumatology : orthopedie traumatologie, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Takatoshi MorookaDepartment of Orthopaedic Surgery, Hyogo College of Medicine, 1-1 Mukogawa-cho, Nishinomiya City, Hyogo, 663-8501, Japan. morooka0818@gmail.com.
Makiko OkunoDepartment of Orthopaedic Surgery, Hyogo College of Medicine, 1-1 Mukogawa-cho, Nishinomiya City, Hyogo, 663-8501, Japan.
Daisuke SeinoDepartment of Orthopaedic Surgery, Hyogo College of Medicine, 1-1 Mukogawa-cho, Nishinomiya City, Hyogo, 663-8501, Japan.
Takuya IsekiDepartment of Orthopaedic Surgery, Hyogo College of Medicine, 1-1 Mukogawa-cho, Nishinomiya City, Hyogo, 663-8501, Japan.
Shigeo FukunishiDepartment of Orthopaedic Surgery, Hyogo College of Medicine, 1-1 Mukogawa-cho, Nishinomiya City, Hyogo, 663-8501, Japan.
Syoji KobashiGraduate School of Engineering, University of Hyogo, 2167 Syosya, Himeji, Hyogo, 671-2280, Japan.
Shinichi YoshiyaDepartment of Orthopaedic Surgery, Hyogo College of Medicine, 1-1 Mukogawa-cho, Nishinomiya City, Hyogo, 663-8501, Japan.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeTo investigate intraoperative kinematics during passive flexion using a surgical navigation system for knees undergoing posterior stabilized (PS) total knee arthroplasty (TKA) with an asymmetric helical post-cam design using navigation system.

methodsIn total, 45 knees with both pre- and postoperative kinematic data available were included in the study. Intraoperative kinematic measurements were performed during the course of surgery using the software incorporated in the navigation system. Measurements were performed at the following two time points: (1) before TKA procedure and (2) after TKA implantation. Among the kinematic parameters studied, anterior/posterior translation and axial rotation during flexion were subjected to the analysis.

resultsBefore surgery, physiologic anterior/posterior translational pattern of the tibia during flexion (rollback of the femur) was found in only 15.6% of the knees. After TKA implantation, postoperative kinematic measurement showed no significant change in the tibial translational during knee flexion. Similarly, with regard to rotation, non-physiologic external tibial rotation in early flexion was observed in the majority of the knees before surgery, and this abnormal kinematic pattern remained after the TKA procedure.

conclusionsThe intraoperative three-dimensional motion analysis using a navigation system showed that the physiologic kinematic pattern (anterior translation and internal rotation of the tibia during flexion) of the knee was distorted in osteoarthritic knees undergoing TKA. The abnormal kinematic pattern before surgery was not fully corrected even after implantation of the PS TKA designed to induce natural knee motion; however, no clear relationship between the intraoperative kinematic pattern and knee flexion angle at one year was demonstrated, and the effect of knee kinematics on postoperative knee function and patient's satisfaction is still unclear.

Indexed as

Knee ProsthesisProsthesis DesignAgedAged, 80 and overArthroplasty, Replacement, KneeBiomechanical PhenomenaFemaleHumansIntraoperative PeriodKnee JointMaleMiddle AgedOsteoarthritis, KneeRotationAnterior–posterior translationInternal–external rotationKnee kinematicsPosterior stabilizedSurgical navigationTotal knee arthroplasty

Identifiers

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.