Clinical commissioning of a new patient positioning system, SyncTraX FX4, for intracranial stereotactic radiotherapy.

Satoshi Tanabe, Osamu Umetsu, Toshikazu Sasage, Satoru Utsunomiya, Ryota Kuwabara, Toshiki Kuribayashi, Hiromasa Takatou, Gen Kawaguchi, Hidefumi Aoyama

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.

0numbers the graph read from it
0cells of the map it votes in
7citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

7 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors.

Satoshi TanabeDepartment of Radiation Oncology, Niigata University Medical and Dental Hospital, Niigata, Japan.
Osamu UmetsuDepartment of Radiation Oncology, Uonuma Kikan Hospital, Niigata, Japan.
Toshikazu SasageDepartment of Radiation Oncology, Uonuma Kikan Hospital, Niigata, Japan.
Satoru UtsunomiyaDepartment of Radiological Technology, Niigata University Graduate School of Health Sciences, Niigata, Japan.
Ryota KuwabaraDepartment of Radiation Oncology, Uonuma Kikan Hospital, Niigata, Japan.
Toshiki KuribayashiDepartment of Radiation Oncology, Uonuma Kikan Hospital, Niigata, Japan.
Hiromasa TakatouDepartment of Radiation Oncology, Uonuma Kikan Hospital, Niigata, Japan.
Gen KawaguchiDepartment of Radiation Oncology, Uonuma Kikan Hospital, Niigata, Japan.
Hidefumi AoyamaDepartment of Radiology and Radiation Oncology, Niigata University Graduate School of Medical and Dental Sciences, Niigata, Japan.

Funding

Japan Society for the Promotion of Science 18K15584Japan Society for the Promotion of Science 18K15626
6 · The paper itself

Abstract

BACKGROUND &

aimsA new real-time tracking radiotherapy (RTRT) system, the SyncTraX FX4 (Shimadzu, Kyoto, Japan), consisting of four X-ray tubes and four ceiling-mounted flat panel detectors (FPDs) combined with a linear accelerator, was installed at Uonuma Kikan Hospital (Niigata, Japan) for the first time worldwide. In addition to RTRT, the SyncTraX FX4 system enables bony structure-based patient verification. Here we provide the first report of this system's clinical commissioning for intracranial stereotactic radiotherapy (SRT). MATERIALS &

methodsA total of five tests were performed for the commissioning: evaluations of (1) the system's image quality; (2) the imaging and treatment coordinate coincidence; and (3) the localization accuracy of cone-beam computed tomography (CBCT) and SyncTraX FX4; (4) the measurement of air kerma; (5) an end-to-end test. RESULTS & DISCUSSION: The tests revealed the following. (1) All image quality evaluation items satisfied each acceptable criterion in all FPDs. (2) The maximum offsets among the centers were ≤0.40 mm in all combinations of the FPD and X-ray tubes (preset). (3) The isocenter localization discrepancies between CBCT and preset #3 in the SyncTraX FX4 system were 0.29 ± 0.084 mm for anterior-posterior, -0.19 ± 0.13 mm for superior-inferior, 0.076 ± 0.11 mm for left-right, -0.11 ± 0.066° for rotation, -0.14 ± 0.064° for pitch, and 0.072±0.058° for roll direction. the Pearson's product-moment correlation coefficient between the two systems was >0.98 in all directions. (4) The mean air kerma value for preset #3 was 0.11 ± 0.0002 mGy in predefined settings (80 kV, 200 mA, 50 msec). (5) For 16 combinations of gantry and couch angles, median offset value in all presets was 0.31 mm (range 0.14-0.57 mm).

conclusionOur results demonstrate a competent performance of the SyncTraX FX4 system in terms of the localization accuracy for intracranial SRT.

Indexed as

Phantoms, ImagingAlgorithmsCone-Beam Computed TomographyHumansImage Processing, Computer-AssistedNeoplasmsParticle AcceleratorsPatient PositioningRadiosurgeryRadiotherapy DosageRadiotherapy, Intensity-ModulatedRadiotherapy Planning, Computer-Assistedcommissioningimage-guided radiotherapy (IGRT)stereotactic radiotherapy