ArticleJournal of applied clinical medical physics2018
A technique for total skin electron therapy (TSET) of an anesthetized pediatric patient.
Article in Journal of applied clinical medical physics, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
5 citing papers in PubMed.
- Modified total skin electron treatment for a paraplegic patient.Journal of applied clinical medical physics · 2025Article
- Patient Positioning and Treatment Techniques in Total Skin Irradiation: A Scoping Review.Cancers · 2025Review
- Patient Dose Analysis Using GafchromicTM EBT3 Film: A Retrospective Study with a Four Dual-Field Technique in Total Skin Electron Therapy.Asian Pacific journal of cancer prevention : APJCP · 2023Article
- Total Skin Treatment with Helical Arc Radiotherapy.International journal of molecular sciences · 2023Review
- In vivo monitoring of total skin electron dose using optically stimulated luminescence dosimeters.Reports of practical oncology and radiotherapy : journal of Greatpoland Cancer Center in Poznan and Polish Society of Radiation OncologyArticle
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
Abstract
purposeTotal skin electron therapy (TSET) is a technique to treat cutaneous lymphomas. While TSET is rarely required in pediatric patients, it poses particular problems for the delivery. It was the aim of the present work to develop a method to deliver TSET to young children requiring anesthetics during treatment.
methodsA customized cradle with a thin window base and Poly(methyl-methacrylate) (PMMA) frame was built and the patient was treated in supine position. Two times six fields of 6 MeV electrons spaced by 60° gantry angles were used without electron applicator and a field size of 36 × 36 cm
resultsA 2-yr-old patient was treated in 12 fractions of 1.5 Gy over 2.4 weeks. Dose to skin was verified daily using thermoluminescence dosimetry and/or radiochromic film. The treatment parameters were adjusted slightly based on in vivo dosimetry resulting in a dose distribution for most of the treatment volume within ±20% of the prescribed dose. Six areas were boosted using conventional electron therapy.
conclusionTSET can be delivered to pediatric patients using a customized couch top on a conventional linear accelerator.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.