Evidence map›Paper›PMID 30223820›Full record

ArticleBMC musculoskeletal disorders2018

Model-based roentgen stereophotogrammetric analysis using elementary geometrical shape models: 10 years results of an uncemented acetabular cup component.

Anne Jacobsen, Frank Seehaus, Yutong Hong, Han Cao, Alexander Schuh, Raimund Forst, Stefan Sesselmann

Abstract read
In one paragraph

Article in BMC musculoskeletal disorders, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed, 1 pooled it
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed, 1 synthesis or guideline pooled it.

  1. Pooled it
  2. Trial
  3. Article
  4. Article
  5. Article
  6. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Anne JacobsenDepartment of Orthopaedic Surgery, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Rathsberger Str. 57, 91054, Erlangen, Germany.
Frank SeehausDepartment of Orthopaedic Surgery, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Rathsberger Str. 57, 91054, Erlangen, Germany.
Yutong HongDepartment of Orthopaedic Surgery, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Rathsberger Str. 57, 91054, Erlangen, Germany.
Han CaoDepartment of Orthopaedic Surgery, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Rathsberger Str. 57, 91054, Erlangen, Germany.
Alexander SchuhMuskuloskelettales Zentrum, Klinikum Neumarkt, Nürnberger Str. 12, 92318, Neumarkt i. d. OPf, Germany.
Raimund ForstDepartment of Orthopaedic Surgery, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Rathsberger Str. 57, 91054, Erlangen, Germany.
Stefan SesselmannDepartment of Orthopaedic Surgery, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Rathsberger Str. 57, 91054, Erlangen, Germany. s.sesselmann@oth-aw.de.ORCID http://orcid.org/0000-0002-6963-0233

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundNon-cemented acetabular cup components demonstrated different clinical performance depending on their surface texture or bearing couple. However, clinical osseointegration needs to be proved for each total joint arthroplasty (TJA) design. Aim of this study was to detect the in vivo migration pattern of a non-cemented cup design, using model-based roentgen stereophotogrammetric analysis with elementary geometrical shape models (EGS-RSA) to calculate early cup migration.

methodsInterchangeable applicability of the model-based EGS-RSA method next to gold standard marker-based RSA method was assessed by clinical radiographs. Afterwards, in vivo acetabular cup migration for 39 patients in a maximum follow up of 120 months (10 years) was calculated using model-based EGS-RSA.

resultsFor the axes with the best predictive capability for acetabular cup loosening, mean (±SD) values were calculated for migration and rotation of the cup. The cup migrated 0.16 (±0.22) mm along the cranio-caudal axis after 24 months and 0.36 (±0.72) mm after 120 months, respectively. It rotated - 0.61 (±0.57) deg. about the medio-lateral axis after 24 months and - 0.53 (±0.67) deg. after 120 months, respectively.

conclusionsInterchangeable applicability of model-based EGS-RSA next to gold standard marker-based RSA method could be shown. Model-based EGS-RSA enables an in vivo migration measurement without the necessity of TJA specific surface models. Migration of the investigated acetabular cup component indicates significant migration values along all the three axes. However, migration values after the second postoperative year were within the thresholds reported in literature, indicating no risk for later aseptic component loosening of this TJA design.

Indexed as

Hip ProsthesisPatient-Specific ModelingProsthesis FailureRadiostereometric AnalysisAcetabulumAdultAgedArthroplasty, Replacement, HipBiomechanical PhenomenaFemaleForeign-Body MigrationHip JointHumansMaleMiddle AgedPredictive Value of TestsTitaniumAcetabular cupElementary geometrical shapesImplant migrationRoentgen stereophotogrammetric analysisRSATotal hip replacement

Identifiers

PMID30223820
PMCPMC6142331

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.