Technical Note: Characterization of clinical linear accelerator triggering latency for motion management system development.
Article in Medical physics, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
7 citing papers in PubMed.
- Article
- Comparison of protons and very high-energy electrons transmission pencil-beam-scanning for FLASH radiotherapy.Physics and imaging in radiation oncology · 2025Article
- Film measurement and analytical approach for assessing treatment accuracy and latency in a magnetic resonance-guided radiotherapy system.Journal of applied clinical medical physics · 2023Article
- Individual pulse monitoring and dose control system for pre-clinical implementation of FLASH-RT.Physics in medicine and biology · 2022Article
- Markerless Pancreatic Tumor Target Localization Enabled By Deep Learning.International journal of radiation oncology, biology, physics · 2019Article
- Deformable abdominal phantom for the validation of real-time image guidance and deformable dose accumulation.Journal of applied clinical medical physics · 2019Article
- Towards real-time respiratory motion prediction based on long short-term memory neural networks.Physics in medicine and biology · 2019Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
Abstract
purposeLatencies for motion management systems have previously been presented as guidelines for system development and implementation. These guidelines consider the overall system latency, including data acquisition, algorithm processing, and linac triggering time. However, during system development, the triggering latency of the clinical linear accelerator is often considered fixed. This paper presents a method to decouple the linac-only triggering latency from the total system latency such that latency can be considered in terms of only the linac-independent aspects of the system.
methodsThe linac-only latency was investigated by considering the time at which a linac response was observed relative to the time at which a beam-on/off triggering signal was sent to the linac. The relative time between the two signals was analyzed using a multichannel oscilloscope with input signals from a custom gating box to manually trigger the beam state as well as a diode positioned at beam isocenter to monitor the linac response. The beam-on/off latency was measured at multiple energies (6/18 MV) and repetition rates (100-600 MU/min) to investigate beam setting dependencies.
resultsThe measured latency was observed to be dependent on the accelerator settings for repetition rate and energy, with beam-on latencies decreasing with increasing repetition rate and decreasing energy. In contrast, the opposite trend was present for the observed beam-off latency. At 600 MU/min, beam-on/off latencies were observed to be 3.37/1.45 ms for a 6 MV beam and 6.02/0.73 ms for an 18 MV beam. Negative latencies were possible for beam-off measurements due to the mechanical latency being less than the pulse separation at given repetition rates.
conclusionsThe linac latency associated with triggering the beam-on/off was determined to have a minor contribution to the total allowable system latency; thus, the majority of the total system latency can be attributed to linac-independent factors.