Evidence map›Paper›PMID 30191344›Full record

ArticleMagma (New York, N.Y.)2018

Simultaneous slice excitation for accelerated passive marker tracking via phase-only cross correlation (POCC) in MR-guided needle interventions.

Andreas Reichert, Michael Bock, Simon Reiss, Christiaan G Overduin, Jurgen J Fütterer, Axel Joachim Krafft

Abstract read
PubMed Publisher
In one paragraph

Article in Magma (New York, N.Y.), 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Andreas ReichertDepartment of Radiology, Medical Physics, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Breisacher Str. 60a, 79106, Freiburg, Germany. andreas.reichert@uniklinik-freiburg.de.ORCID http://orcid.org/0000-0002-0656-4144
Michael BockDepartment of Radiology, Medical Physics, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Breisacher Str. 60a, 79106, Freiburg, Germany.
Simon ReissDepartment of Radiology, Medical Physics, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Breisacher Str. 60a, 79106, Freiburg, Germany.
Christiaan G OverduinDepartment of Radiology and Nuclear Medicine, Radboud University Medical Center, Nijmegen, The Netherlands.
Jurgen J FüttererDepartment of Radiology and Nuclear Medicine, Radboud University Medical Center, Nijmegen, The Netherlands.
Axel Joachim KrafftDepartment of Radiology, Medical Physics, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Breisacher Str. 60a, 79106, Freiburg, Germany.

Funding

Deutsche Forschungsgemeinschaft HA 7006/1-1
6 · The paper itself

Abstract

objectiveTo accelerate a passive tracking sequence based on phase-only cross correlation (POCC) using simultaneous slice excitation.

methodsFor magnetic resonance (MR)-guided biopsy procedures, passive markers have been proposed that can be automatically localized online using a POCC-based tracking sequence. To accelerate the sequence, a phase-offset multiplanar (POMP) excitation technique was implemented to acquire tracking images. In a phantom experiment, the POMP-POCC sequence was tested and compared with the sequential non-accelerated version in terms of duration and accuracy. Further, technical feasibility of the POMP-POCC sequence was tested in a patient undergoing a prostate biopsy.

resultsThe temporal resolution of the POMP-POCC tracking sequence is accelerated by 33% compared with the sequential approach. In phantom experiments, the POMP-POCC and sequential sequences yielded the same targeting accuracy of 1.6 ± 0.7 mm. Technical proof of concept of the new sequence could be demonstrated in a successful in vivo prostate biopsy.

conclusionPOMP-POCC tracking can substantially reduce the duration of localization of passive markers in MR-guided needle interventions without compromising targeting accuracy.

Indexed as

Magnetic Resonance Imaging, InterventionalNeedlesAlgorithmsArtifactsBiopsyEquipment DesignHumansImage Processing, Computer-AssistedMalePhantoms, ImagingProstateTime FactorsInterventional magnetic resonance imagingMagnetic-resonance-guided interventional proceduresMR-guided interventional procedures

Identifiers

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked