Image quality and dose characteristics for an O-arm intraoperative imaging system with model-based image reconstruction.
Article in Medical physics, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 9 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
9 citing papers in PubMed.
- Article
- Intraoperative cone-beam and slot-beam CT: 3D image quality and dose with a slot collimator on the O-arm imaging system.Medical physics · 2021Article
- Three-dimensional intraoperative computed tomography imaging for zygomatic fracture repair.Journal of the Korean Association of Oral and Maxillofacial Surgeons · 2021Article
- Hip capsular strain varies between ligaments dependent on both hip position- and applied rotational force.Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA · 2020Article
- A mobile isocentric C-arm for intraoperative cone-beam CT: Technical assessment of dose and 3D imaging performance.Medical physics · 2020Article
- Known-component metal artifact reduction (KC-MAR) for cone-beam CT.Physics in medicine and biology · 2019Article
- Known-component 3D image reconstruction for improved intraoperative imaging in spine surgery: A clinical pilot study.Medical physics · 2019Article
- Radiation dose and image quality comparison during spine surgery with two different, intraoperative 3D imaging navigation systems.Journal of applied clinical medical physics · 2019Article
- Cone-Beam CT image contrast and attenuation-map linearity improvement (CALI) for brain stereotactic radiosurgery procedures.Journal of applied clinical medical physics · 2018Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
Abstract
purposeTo assess the imaging performance and radiation dose characteristics of the O-arm CBCT imaging system (Medtronic Inc., Littleton MA) and demonstrate the potential for improved image quality and reduced dose via model-based image reconstruction (MBIR).
methodsTwo main studies were performed to investigate previously unreported characteristics of the O-arm system. First is an investigation of dose and 3D image quality achieved with filtered back-projection (FBP) - including enhancements in geometric calibration, handling of lateral truncation and detector saturation, and incorporation of an isotropic apodization filter. Second is implementation of an MBIR algorithm based on Huber-penalized likelihood estimation (PLH) and investigation of image quality improvement at reduced dose. Each study involved measurements in quantitative phantoms as a basis for analysis of contrast-to-noise ratio and spatial resolution as well as imaging of a human cadaver to test the findings under realistic imaging conditions.
resultsView-dependent calibration of system geometry improved the accuracy of reconstruction as quantified by the full-width at half maximum of the point-spread function - from 0.80 to 0.65 mm - and yielded subtle but perceptible improvement in high-contrast detail of bone (e.g., temporal bone). Standard technique protocols for the head and body imparted absorbed dose of 16 and 18 mGy, respectively. For low-to-medium contrast (<100 HU) imaging at fixed spatial resolution (1.3 mm edge-spread function) and fixed dose (6.7 mGy), PLH improved CNR over FBP by +48% in the head and +35% in the body. Evaluation at different dose levels demonstrated 30% increase in CNR at 62% of the dose in the head and 90% increase in CNR at 50% dose in the body.
conclusionsA variety of improvements in FBP implementation (geometric calibration, truncation and saturation effects, and isotropic apodization) offer the potential for improved image quality and reduced radiation dose on the O-arm system. Further gains are possible with MBIR, including improved soft-tissue visualization, low-dose imaging protocols, and extension to methods that naturally incorporate prior information of patient anatomy and/or surgical instrumentation.