Evidence map›Paper›PMID 30054714›Full record

ArticleHerz2020

What physicians do in case of a failure of the pace-sense part of a defibrillation lead : Survey in Germany, Austria and Switzerland.

P Lacour, A Parwani, M Huemer, P Attanasio, P L Dang, J Luebcke, L Schleussner, D Blaschke, L-H Boldt, B Pieske and 2 more

Abstract read
PubMed Publisher
In one paragraph

Article in Herz, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

12 authors.

P LacourDepartment of Cardiology, Charité-Universitaetsmedizin Berlin, Campus Virchow-Klinikum, Augustenburger Platz 1, 13353, Berlin, Germany.
A ParwaniDepartment of Cardiology, Charité-Universitaetsmedizin Berlin, Campus Virchow-Klinikum, Augustenburger Platz 1, 13353, Berlin, Germany.
M HuemerDepartment of Cardiology, Charité-Universitaetsmedizin Berlin, Campus Virchow-Klinikum, Augustenburger Platz 1, 13353, Berlin, Germany.
P AttanasioDepartment of Cardiology, Charité-Universitaetsmedizin Berlin, Campus Virchow-Klinikum, Augustenburger Platz 1, 13353, Berlin, Germany.
P L DangDepartment of Cardiology, Charité-Universitaetsmedizin Berlin, Campus Virchow-Klinikum, Augustenburger Platz 1, 13353, Berlin, Germany.
J LuebckeDepartment of Cardiology, Charité-Universitaetsmedizin Berlin, Campus Virchow-Klinikum, Augustenburger Platz 1, 13353, Berlin, Germany.
L SchleussnerDepartment of Cardiology, Charité-Universitaetsmedizin Berlin, Campus Virchow-Klinikum, Augustenburger Platz 1, 13353, Berlin, Germany.
D BlaschkeDepartment of Cardiology, Charité-Universitaetsmedizin Berlin, Campus Virchow-Klinikum, Augustenburger Platz 1, 13353, Berlin, Germany.
L-H BoldtDepartment of Cardiology, Charité-Universitaetsmedizin Berlin, Campus Virchow-Klinikum, Augustenburger Platz 1, 13353, Berlin, Germany.
B PieskeDepartment of Cardiology, Charité-Universitaetsmedizin Berlin, Campus Virchow-Klinikum, Augustenburger Platz 1, 13353, Berlin, Germany.
W HaverkampDepartment of Cardiology, Charité-Universitaetsmedizin Berlin, Campus Virchow-Klinikum, Augustenburger Platz 1, 13353, Berlin, Germany.
F BlaschkeDepartment of Cardiology, Charité-Universitaetsmedizin Berlin, Campus Virchow-Klinikum, Augustenburger Platz 1, 13353, Berlin, Germany. florian.blaschke@charite.de.ORCID http://orcid.org/0000-0001-7193-0318

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundThe possible treatment strategies for defects of the pace-sense (P/S) part of a defibrillation lead are either implantation of a new high-voltage (HV)-P/S lead, with or without extraction of the malfunctioning lead, or implantation of a P/S lead.

methodsWe conducted a Web-based survey across cardiac implantable electronic device (CIED) centers to investigate their procedural practice and decision-making process in cases of failure of the P/S portion of defibrillation leads. In particular, we focused on the question of whether the integrity of the HV circuit is confirmed by a test shock before decision-making. The questionnaire included 14 questions and was sent to 951 German, 341 Austrian, and 120 Swiss centers.

resultsThe survey was completed by 183 of the 1412 centers surveyed (12.7% response rate). Most centers (90.2%) do not conduct a test shock to confirm the integrity of the HV circuit before decision-making. Procedural practice in lead management varies depending on the presentation of lead failure and whether the center applies a test shock. In centers that do not conduct a test shock, the majority (69.9%) implant a new HV-P/S lead. Most centers (61.7%) that test the integrity of the HV system implant a P/S lead. The majority of centers favor DF-4 connectors (74.1%) over DF-1 connectors (25.9%) at first CIED implantation.

conclusionEither implanting a new HV-P/S lead or placing an additional P/S lead are selected strategies if the implantable cardioverter-defibrillator lead failure is localized to the P/S portion. However, conducting a test shock to confirm the integrity of the HV component is rarely performed.

Indexed as

Defibrillators, ImplantableElectric CountershockPractice Patterns, Physicians'AustriaGermanySurveys and QuestionnairesSwitzerlandCardiac implantable electronic deviceCardiac resynchronization therapyHigh-voltage leadImplantable cardioverter-defibrillatorMedical device failure

Identifiers

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked