Evidence map›Paper›PMID 29961040›Full record

ArticleBMJ open2018

Does the architectural layout of a NICU affect alarm pressure? A comparative clinical audit of a single-family room and an open bay area NICU using a retrospective study design.

Rohan Joshi, Henrica van Straaten, Heidi van de Mortel, Xi Long, Peter Andriessen, Carola van Pul

Abstract readComparative Study
In one paragraph

Article in BMJ open, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
7citing papers in PubMed, 1 pooled it
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

7 citing papers in PubMed, 1 synthesis or guideline pooled it.

  1. Pooled it
  2. Article
  3. Observational
  4. Article
  5. Article
  6. Article
  7. Predictive Monitoring of Critical Cardiorespiratory Alarms in Neonates Under Intensive Care.IEEE journal of translational engineering in health and medicine · 2019
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Rohan JoshiDepartment of Industrial Design, Eindhoven University of Technology, Eindhoven, The Netherlands.ORCID 0000-0001-6504-9399
Henrica van StraatenDepartment of Neonatology, Isala, Zwolle, The Netherlands.
Heidi van de MortelDepartment of Neonatology, Maxima Medisch Centrum locatie Veldhoven, Veldhoven, The Netherlands.
Xi LongDepartment of Fertility, Pregnancy, and Parenting Solutions, Philips Research Eindhoven, Eindhoven, The Netherlands.ORCID 0000-0001-9505-1270
Peter AndriessenDepartment of Neonatology, Maxima Medisch Centrum locatie Veldhoven, Veldhoven, The Netherlands.
Carola van PulDepartment of Clinical Physics, Maxima Medisch Centrum locatie Veldhoven, Veldhoven, Noord-Brabant, The Netherlands.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectivesTo determine differences in alarm pressure between two otherwise comparable neonatal intensive care units (NICUs) differing in architectural layout-one of a single-family room (SFR) design and the other of an open bay area (OBA) design.

designRetrospective audit of more than 2000 patient days from each NICU cataloguing the differences in the number and duration of alarms for critical and alerting alarms, as well as the interaction of clinicians with the patient monitor.

settingTwo level 3 NICUs.

resultsA total of more than 150 000 critical and 1.2 million alerting alarms were acquired from the two NICUs. The number of audible alarms and the associated noise pollution varied considerably with the OBA NICU generating 44% more alarms per infant per day even though the SFR NICU generated 2.5 as many critical desaturation alarms per infant per day.

conclusionDifferences in the architectural layout of NICUs and the consequent differences in delays, thresholds and distribution systems for alarms are associated with differences in alarm pressure.

Indexed as

Clinical AuditIntensive Care Units, NeonatalPatients' RoomsClinical AlarmsHospital Design and ConstructionHumansInfantInfant, NewbornInfant, PrematureMonitoring, PhysiologicRetrospective Studiesclinical auditneonatology