Evidence map›Paper›PMID 29953485›Full record

ArticlePloS one2018

Effect of pannus formation on the prosthetic heart valve: In vitro demonstration using particle image velocimetry.

Hojin Ha, Hyun Jung Koo, Hyung Kyu Huh, Guk Bae Kim, Jihoon Kweon, Namkug Kim, Young-Hak Kim, Joon-Won Kang, Tae-Hwan Lim, Jae-Kwan Song and 2 more

Open access · goldAbstract read
In one paragraph

Article in PloS one, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed, 1 pooled it
1.9field-weighted citation impact, top 13% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed, 1 synthesis or guideline pooled it, 17 citations in OpenAlex.

  1. Pooled it
  2. Role of Cardiac CT in the Evaluation of Valvular Heart Disease.Journal of the Korean Society of Radiology · 2026
    Review
  3. The Role of Multimodality Imaging in Left-Sided Prosthetic Valve Dysfunction.Journal of cardiovascular development and disease · 2022
    Review
  4. Mechanical prosthetic valve thrombosis in current era: 5-year follow-up.Indian journal of thoracic and cardiovascular surgery · 2021
    Article
  5. Article
  6. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

12 authors at 3 institutions in 1 country.

Hojin HaDepartment of Mechanical and Biomedical Engineering, Kangwon National University, Chuncheon, South Korea.
Hyun Jung KooDepartment of Radiology, University of Ulsan College of Medicine, Asan Medical Center, Seoul, South Korea.
Hyung Kyu HuhDepartment of Mechanical Engineering, Pohang University of Science and Technology, Pohang, South Korea.
Guk Bae KimAsan Institute of Life Science, Asan Medical Center, University of Ulsan College of Medicine, Seoul, South Korea.
Jihoon KweonDepartment of Cardiology, University of Ulsan College of Medicine, Asan Medical Center, Seoul, South Korea.
Namkug KimDepartment of Convergence Medicine, University of Ulsan College of Medicine, Asan Medical Center, Seoul, South Korea.
Young-Hak KimDepartment of Cardiology, University of Ulsan College of Medicine, Asan Medical Center, Seoul, South Korea.
Joon-Won KangDepartment of Radiology, University of Ulsan College of Medicine, Asan Medical Center, Seoul, South Korea.
Tae-Hwan LimDepartment of Radiology, University of Ulsan College of Medicine, Asan Medical Center, Seoul, South Korea.
Jae-Kwan SongDepartment of Cardiology, University of Ulsan College of Medicine, Asan Medical Center, Seoul, South Korea.
Sang Joon LeeDepartment of Mechanical Engineering, Pohang University of Science and Technology, Pohang, South Korea.
Dong Hyun YangDepartment of Radiology, University of Ulsan College of Medicine, Asan Medical Center, Seoul, South Korea.ORCID 0000-0001-5477-558X
Ulsan College · KRPohang University of Science and Technology · KRKangwon National University · KR

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Although hemodynamic influence of the subprosthetic tissue, termed as pannus, may contribute to prosthetic aortic valve dysfunction, the relationship between pannus extent and hemodynamics in the prosthetic valve has rarely been reported. We investigated the fluid dynamics of pannus formation using in vitro experiments with particle image velocimetry. Subvalvular pannus formation caused substantial changes in prosthetic valve transvalvular peak velocity, transvalvular pressure gradient (TPG) and opening angle. Maximum flow velocity and corresponding TPG were mostly affected by pannus width. When the pannus width was 25% of the valve diameter, pannus formation elevated TPG to >2.5 times higher than that without pannus formation. Opening dysfunction was observed only for a pannus involvement angle of 360°. Although circumferential pannus with an involvement angle of 360° decreased the opening angle of the valve from approximately 82° to 58°, eccentric pannus with an involvement angle of 180° did not induce valve opening dysfunction. The pannus involvement angle largely influenced the velocity flow field at the aortic sinus and corresponding hemodynamic indices, including wall shear stress, principal shear stress and viscous energy loss distributions. Substantial discrepancy between the velocity-based TPG estimation and direct pressure measurements was observed for prosthetic valve flow with pannus formation.

Indexed as

Heart Valve ProsthesisHemodynamicsModels, CardiovascularFemaleHumansMaleRetrospective Studies

Identifiers

PMID29953485
PMCPMC6023143
OpenAlexW2810169966

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked