Inductively Coupled Plasma Mass Spectrometry as a Reference Method to Evaluate Serum Calcium Measurement Bias and the Commutability of Processed Materials during Routine Measurements.
Article in Chinese medical journal, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- Hypoparathyroidism: a brief historical overview for clinicians.Frontiers in endocrinology · 2026Review
- [Significance and methodology of monitoring calcemia in disorders of mineral metabolism: challenges and prospects].Problemy endokrinologii · 2024Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundMeasuring total serum calcium is important for the diagnosis of diseases. Currently, results from commercial kits for calcium measurement are variable. Generally, the performance of serum calcium measurements is monitored by external quality assessment (EQA) or proficiency testing schemes. However, the commutability of the EQA samples and calibrators is often unknown, which limits the effectiveness of EQA schemes. The aim of this study was to evaluate the bias of serum calcium measurements and the commutability of processed materials.
methodsInductively coupled plasma mass spectrometry was applied as a comparative method, and 14 routine methods were chosen as test methods. Forty-eight serum samples from individual patients and 25 processed materials were quantified. A scatter plot was generated from patient samples, and 95% prediction intervals were calculated to evaluate the commutability of the processed materials and measurement bias at three concentration levels was used to determine the accuracy of routine assays.
resultsAll assays showed high precision (total coefficient of variation [CV] <2.26%) and correlation coefficients (r > 0.99). For all assays, the mean bias for the 48 patient samples ranged from -0.13 mmol/L to 0.00 mmol/L (-5.61-0.01%), and the ranges for the three concentrations were -0.10-0.04 mmol/L (-5.71-2.35%), -0.14--0.01 mmol/L (-5.80--0.30%), and -0.19-0.04 mmol/L (-6.24-1.22%). The EQA samples, calibrators, and animal sera exhibited matrix effects in some assays; human serum pools were commutable in all assays; certificate reference materials were commutable in most assays, and only GBW09152 exhibited a matrix effect in one assay; and aqueous reference materials exhibited matrix effects in most assays.
conclusionsBiases for most assays were within the acceptable range, although the accuracy of some assays needs improvement. Human serum pools prepared from patient samples were commutable, and the other tested materials exhibited a matrix effect.