ArticleJournal of medical signals and sensors
Evaluation of Targeted Image-Guided Radiation Therapy Treatment Planning System by Use of American Association of Physicists in Medicine Task Group-119 Test Cases.
Article in Journal of medical signals and sensors. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundThis study aimed to evaluate the overall accuracy of the beam commissioning criteria of targeted image-guided radiation therapy (TiGRT) treatment planning system (TPS) based on the American Association of Physicists in Medicine (AAPM) Task Group Report 119 (TG-119).
methodsThe work was performed using 6 MV energy LINAC with a variable dose rate of 200 MU/min which equipped with the high-quality external TiGRT dynamic multileaf collimator model H. The AAPM TG-119 intensity-modulated radiation therapy (IMRT) commissioning tests are composed of two preliminary tests and four clinical test cases. The clinical tests consisted of mock prostate, mock head and neck, C-shaped target, and multitarget. EDR2 film was used for evaluating the IMRT plans and point dose measured by a Pinpoint chamber positioned in slab phantom. The film analysis was done with the Sun Nuclear Corporation patient software. The dose prescription for each fraction was 200 cGy in mock prostate, mock head and neck, C-shaped target, and multitarget. Dose distributions were analyzed using gamma criteria of 3% and 2% dose difference (DD) and 3 and 2 mm distance to agreement.
resultsIn all test cases, the gamma criteria for 2%/2 and 3%/3 were found to be 94% and 98%, respectively. Results showed that the average gamma criteria result was in the range of 99.1% to 93% (3%/3, 2%/2) overall test cases.
conclusionsFindings were favorable and in some tests were comparable with the other studies. The dose point values were within the mean values of the range reported by TG-119. Overall, the TiGRT TPS is needed to apply IMRT technique in radiation therapy centers.