ArticleJournal of diabetes science and technology2018
Evaluation of Four HbA1c Point-of-Care Devices Using International Quality Targets: Are They Fit for the Purpose?
Article in Journal of diabetes science and technology, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 11 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
11 citing papers in PubMed.
- Observational
- Evaluation of a point-of-care testing platform against a laboratory HbA1c measurement method in a South African tertiary-level diabetes clinic.Practical laboratory medicine · 2026Article
- Point-of-care HbA1c testing in a tertiary, university referral center: living up to the potential?Endocrine · 2026Observational
- Glycated Haemoglobin (HbAJournal of clinical medicine · 2024Review
- Trends in Quantification of HbA1c Using Electrochemical and Point-of-Care Analyzers.Sensors (Basel, Switzerland) · 2023Review
- Current Challenges and Future Trends of Enzymatic Paper-Based Point-of-Care Testing for Diabetes Mellitus Type 2.Biosensors · 2021Review
- Hemoglobin ADiabetes care · 2021Review
- Exposure to cholinesterase inhibiting insecticides and blood glucose level in a population of Ugandan smallholder farmers.Occupational and environmental medicine · 2020Article
- Fingerstick Precision and Total Error of a Point-of-Care HbA1c Test.Journal of diabetes science and technology · 2020Article
- Analysis: Investigating the quality of POCT devices for HbA1c, what are our next steps?Journal of diabetes science and technology · 2019Article
- Verification of the HbA1c method on the STANDARD F2400Nigerian medical journal : journal of the Nigeria Medical AssociationArticle
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
2 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundPoint-of-care (POC) testing is becoming increasingly valuable in health care delivery, and it is important that the devices used meet the same quality criteria as main laboratory analyzers. While external quality assessment (EQA) provides a great tool for assessing quality, many POC devices are not enrolled in these schemes and standard laboratory evaluations are needed to assess performance.
methodsThe Clinical and Laboratory Standards Institute (CLSI) protocols EP-5 and EP-9 were applied to investigate imprecision, accuracy and bias. We assessed bias using the mean of 4 certified secondary reference measurement procedures (SRMPs).
resultsThe Afinion2™ and the Quo-Lab had CVs of ≤1.7 and ≤2.4% respectively in IFCC SI units (≤1.2 and ≤1.7% NGSP) and a bias ≤2 mmol/mol (≤0.2% NGSP) at 48 and 75 mmol/mol (6.5 and 9.0% NGSP). Sigma for the Afinion2 was 5.8 and for the Quo-Lab 4.0. Both methods passed the NGSP criteria with 2 instruments when compared with 4 individual SRMPs. The HbA1c 501 had a CV of 3.4% and 2.7% in IFCC SI units (2.1% and 1.7% NGSP) and a bias ≤2.4 mmol/mol (≤0.2% NGSP) and passed the NGSP criteria with 2 instruments compared with 4 individual SRMPs except for instrument 2 compared with the Tosoh G8. Sigma was 2.1. The A1Care had a sigma of 1.4 and failed all criteria mainly due to a high CV (6.2% and 4.1% in IFCC SI units [4.1% and 2.9% NGSP] at 48 and 75 mmol/mol [6.5 and 9.0% NGSP]).
conclusionsThe analytical performance was excellent for the Afinion2 and the Quo-Lab, acceptable for the HbA1c 501 and unacceptable for the A1Care according to different used criteria, demonstrating that whilst performance is improving there are still areas for considerable improvement.