Evidence map›Paper›PMID 29903779›Full record

ArticleBMJ case reports2018

Alarming increase in HbA1c and near misdiagnosis of diabetes mellitus resulting from a clinical laboratory instrument upgrade and haemoglobin variant.

Steven D Chessler, Donald E Lee

Open access · bronzeAbstract readCase Reports
In one paragraph

Article in BMJ case reports, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
0.3field-weighted citation impact, top 37% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed, 3 citations in OpenAlex.

  1. Article
  2. Hemoglobin Wayne causing a falsely elevated hemoglobin A1c.Proceedings (Baylor University. Medical Center) · 2022
    Article
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

2 authors at 1 institution in 1 country.

Steven D ChesslerDepartment of Medicine, University of California Irvine, Irvine, California, USA.ORCID http://orcid.org/0000-0002-0212-546X
Donald E LeeDepartment of Medicine, University of California Irvine, Irvine, California, USA.
University of California, Irvine · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

A 29-year-old woman was referred for new-onset diabetes mellitus after her glycosylated haemoglobin (HbA1c) was found to be 10.2%. Three years earlier, the patient's HbA1c-measured by the same clinical laboratory-had been 5.5%. The newer HbA1c was discordant with fasting glucose levels and a lack of diabetes-associated symptoms. The laboratory reported that their assay methodology remained unchanged and also that no haemoglobin variants were detected. Further investigation, however, revealed, first, that the patient carried a haemoglobin alpha chain mutation (Hb Wayne) that can sometimes cause assay interference and, second, that although the laboratory's assay methodology had not changed, their assay instrument had. Depending on assay methodology, haemoglobin variants can cause HbA1c assay interference and the presence of these variants may not be detected by the performing laboratory. Interference may not only be dependent on assay methodology but also on the assay instrument used.

Indexed as

Diagnostic ErrorsAdultCarrier StateDiabetes Mellitus, Type 2FemaleGlycated HemoglobinHematologic TestsHemoglobins, AbnormalHumansPatient Outcome AssessmentGlycated HemoglobinHemoglobins, Abnormalhemoglobin Waynediabetesendocrinologyhaematology (incl blood transfusion)

Identifiers

PMID29903779
PMCPMC6011579
OpenAlexW2808477076

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked