ArticleBMJ case reports2018
Alarming increase in HbA1c and near misdiagnosis of diabetes mellitus resulting from a clinical laboratory instrument upgrade and haemoglobin variant.
Article in BMJ case reports, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed, 3 citations in OpenAlex.
- The Hemoglobin Wayne Variant and Association With Falsely Elevated HbAJCEM case reports · 2023Article
- Hemoglobin Wayne causing a falsely elevated hemoglobin A1c.Proceedings (Baylor University. Medical Center) · 2022Article
- Glycosylated haemoglobin: a false sense of security.BMJ case reports · 2018Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
2 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
A 29-year-old woman was referred for new-onset diabetes mellitus after her glycosylated haemoglobin (HbA1c) was found to be 10.2%. Three years earlier, the patient's HbA1c-measured by the same clinical laboratory-had been 5.5%. The newer HbA1c was discordant with fasting glucose levels and a lack of diabetes-associated symptoms. The laboratory reported that their assay methodology remained unchanged and also that no haemoglobin variants were detected. Further investigation, however, revealed, first, that the patient carried a haemoglobin alpha chain mutation (Hb Wayne) that can sometimes cause assay interference and, second, that although the laboratory's assay methodology had not changed, their assay instrument had. Depending on assay methodology, haemoglobin variants can cause HbA1c assay interference and the presence of these variants may not be detected by the performing laboratory. Interference may not only be dependent on assay methodology but also on the assay instrument used.