Evidence map›Paper›PMID 29786775›Full record

ArticleAnnals of biomedical engineering2018

Stormram 4: An MR Safe Robotic System for Breast Biopsy.

Vincent Groenhuis, Françoise J Siepel, Jeroen Veltman, Jordy K van Zandwijk, Stefano Stramigioli

Open access · hybridAbstract read
In one paragraph

Article in Annals of biomedical engineering, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.

0numbers the graph read from it
0cells of the map it votes in
8citing papers in PubMed
4.5field-weighted citation impact, top 5% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

8 citing papers in PubMed, 52 citations in OpenAlex.

  1. Low-Complexity Robotic Device for Magnetic Resonance Imaging-Guided Needle Biopsy.The international journal of medical robotics + computer assisted surgery : MRCAS · 2025
    Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Simple methods to test the accuracy of MRgFUS robotic systems.The international journal of medical robotics + computer assisted surgery : MRCAS · 2021
    Article
  7. Article
  8. Artificial intelligence in dermatology and healthcare: An overview.Indian journal of dermatology, venereology and leprology
    Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 1 institution in 1 country.

Vincent GroenhuisDepartment of Robotics and Mechatronics, University of Twente, att. Vincent Groenhuis, Room CR-3.526, Postbus 217, 7500AE, Enschede, The Netherlands. v.groenhuis@utwente.nl.ORCID http://orcid.org/0000-0002-1073-6690
Françoise J SiepelDepartment of Robotics and Mechatronics, University of Twente, att. Vincent Groenhuis, Room CR-3.526, Postbus 217, 7500AE, Enschede, The Netherlands.
Jeroen VeltmanDepartment of Robotics and Mechatronics, University of Twente, att. Vincent Groenhuis, Room CR-3.526, Postbus 217, 7500AE, Enschede, The Netherlands.
Jordy K van ZandwijkDepartment of Robotics and Mechatronics, University of Twente, att. Vincent Groenhuis, Room CR-3.526, Postbus 217, 7500AE, Enschede, The Netherlands.
Stefano StramigioliDepartment of Robotics and Mechatronics, University of Twente, att. Vincent Groenhuis, Room CR-3.526, Postbus 217, 7500AE, Enschede, The Netherlands.
University of Twente · NL

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Suspicious lesions in the breast that are only visible on magnetic resonance imaging (MRI) need to be biopsied under MR guidance with high accuracy and efficiency for accurate diagnosis. The aim of this study is to present a novel robotic system, the Stormram 4, and to perform preclinical tests in an MRI environment. Excluding racks and needle, its dimensions are 72 × 51 × 40 mm. The Stormram 4 is driven by two linear and two curved pneumatic stepper motors. The linear motor is capable of exerting 63 N of force at a pressure of 0.65 MPa. In an MRI environment the maximum observed stepping frequency is 30 Hz (unloaded), or 8 Hz when full force is needed. The Stormram 4's mean positioning error is 0.73 ± 0.47 mm in free air, and 1.29 ± 0.59 mm when targeting breast phantoms in MRI. Excluding the off-the-shelf needle, the robot is inherently MR safe. The robot is able to accurately target lesions under MRI guidance, reducing tissue damage and risk of false negatives. These results are promising for clinical experiments, improving the quality of healthcare in the field of MRI-guided breast biopsies.

Indexed as

Robotic Surgical ProceduresBiopsy, Fine-NeedleBreastFemaleHumansMagnetic resonance imagingPneumaticsRapid prototypingStepper motor

Identifiers

PMID29786775
PMCPMC6153978
OpenAlexW2804900779

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.