Comparison of hemodialysis arteriovenous fistula blood flow rates measured by Doppler ultrasound and phase-contrast magnetic resonance imaging.
Article in Journal of vascular surgery, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 12 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
12 citing papers in PubMed, 25 citations in OpenAlex.
- Predicting haemodialysis arteriovenous fistula outcomes using computational fluid dynamics and ferumoxytol-enhanced MRI.The journal of vascular access · 2026Article
- Assessment of volume flow rate in arteriovenous fistulas with a novel ultrasound Doppler device (earlybird): Trend analysis, comparison of methods, and inter- and intra-rater reliability.The journal of vascular access · 2025Article
- Hemodynamics are associated with subsequent lumen remodeling and clinical maturation of hemodialysis arteriovenous fistula.Scientific reports · 2025Article
- Review
- Association of Shear Stress with Subsequent Lumen Remodeling in Hemodialysis Arteriovenous Fistulas.Clinical journal of the American Society of Nephrology : CJASN · 2023Article
- Influence of Rigid-Elastic Artery Wall of Carotid and Coronary Stenosis on Hemodynamics.Bioengineering (Basel, Switzerland) · 2022Article
- Increased Excess Pressure After Creation of an Arteriovenous Fistula in End-Stage Renal Disease.American journal of hypertension · 2022Article
- Differential hemodynamics between arteriovenous fistulas with or without intervention before successful use.Frontiers in cardiovascular medicine · 2022Article
- Medical Image-Based Computational Fluid Dynamics and Fluid-Structure Interaction Analysis in Vascular Diseases.Frontiers in bioengineering and biotechnology · 2022Review
- Current role of ultrasound in hemodialysis access evaluation.The journal of vascular access · 2021Review
- Analyses of hemodialysis arteriovenous fistula geometric configuration and its associations with maturation and reintervention.Journal of vascular surgery · 2021Article
- A mouse model of stenosis distal to an arteriovenous fistula recapitulates human central venous stenosis.JVS-vascular science · 2020Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors at 4 institutions in 2 countries.
Funding
Abstract
objectiveThe objective of this study was to compare blood flow rates measured by Doppler ultrasound (DUS) and phase-contrast magnetic resonance imaging (MRI) in patients having a hemodialysis arteriovenous fistula (AVF) and to identify scenarios in which there was significant discordance between these two approaches.
methodsBlood flow rates in the proximal artery (PA) and draining vein (DV) of newly created upper extremity AVFs were measured and compared using DUS and phase-contrast MRI at 1 day, 6 weeks, and 6 months postoperatively.
resultsBlood flow rates in the PA measured by DUS (1155 ± 907 mL/min, mean ± standard deviation) and by MRI (1170 ± 657 mL/min) were not statistically different (P = .812) based on 78 data pairs from 49 patients. DV DUS flow (1277 ± 995 mL/min) and MRI flow (1130 ± 655 mL/min) were also not statistically different (P = .071) based on 64 data pairs. In both PA and DV, the two methods substantially agreed with each other (Cohen κ: PA, 0.66; DV, 0.67) when flow rates were put into four clinically relevant categories (<300, 300-599, 600-1499, and ≥1500 mL/min). The Bland-Altman analyses of DUS and MRI flow identified six and four outliers for PA and DV, respectively. Seven outliers had higher DUS than MRI flow, with all DUS scan sites having a large lumen or significant local curvature; the other three had lower DUS flow, partly due to an underestimation of lumen diameter by DUS.
conclusionsDUS and MRI flow rates are generally comparable in both PA and DV. When DUS is used for flow measurements, careful attention to accurate lumen diameter measurements is needed and scan sites with marked curvature should be avoided. Our result may improve the accuracy of DUS-measured AVF blood flow rate.