Assessment of the renal collecting system using a pocket-sized ultrasound device.

Toru Kameda, Kumiko Uebayashi, Kazuko Wagai, Fukiko Kawai, Nobuyuki Taniguchi

Abstract read
PubMed Publisher
In one paragraph

Article in Journal of medical ultrasonics (2001), 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed.

  1. Trial
  2. Accuracy of portable ultrasound machines for obstetric biometry.Ultrasound in obstetrics & gynecology : the official journal of the International Society of Ultrasound in Obstetrics and Gynecology · 2024
    Observational
  3. Article
  4. Article
  5. The Use of Handheld Ultrasound Devices in Emergency Medicine.Current emergency and hospital medicine reports · 2021
    Review
  6. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Toru KamedaDepartment of Clinical Laboratory Medicine, Saiseikai Utsunomiya Hospital, 911-1 Takebayashi, Utsunomiya, Tochigi, 321-0974, Japan. kamekame@pb3.so-net.ne.jp.ORCID http://orcid.org/0000-0002-7538-3427
Kumiko UebayashiDepartment of Clinical Laboratory Medicine, Saiseikai Utsunomiya Hospital, 911-1 Takebayashi, Utsunomiya, Tochigi, 321-0974, Japan.
Kazuko WagaiDepartment of Clinical Laboratory, Saiseikai Utsunomiya Hospital, Utsunomiya, Japan.
Fukiko KawaiDepartment of Clinical Laboratory Medicine, Saiseikai Utsunomiya Hospital, 911-1 Takebayashi, Utsunomiya, Tochigi, 321-0974, Japan.
Nobuyuki TaniguchiDepartment of Clinical Laboratory Medicine, Jichi Medical University, Shimotsuke, Japan.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeTo assess the performance of a pocket-sized ultrasound device (PUD) for evaluating dilatation of the renal collecting system with high-end ultrasound devices (HUDs) as a reference standard.

methodsOne sonographer examined both kidneys using a PUD to evaluate dilatation of the collecting system. The grading of the dilatation ranged from 0 to 4. Immediately after the examination, another sonographer blinded to the previous results performed a formal examination with a HUD.

resultsTwo hundred kidneys in 100 patients were included in the analysis. The agreement of grades between the PUD and HUDs was excellent (weighted kappa = 0.83; P < 0.001). When hydronephrosis was defined as grade 1 or higher, the test characteristics of the PUD were as follows: sensitivity 91% (95% confidence interval (CI) 79-97%), positive predictive value 73% (95% CI 60-83%), and negative predictive value 96% (95% CI 92-99%). When hydronephrosis was defined as grade 2 or higher, the test characteristics were as follows: sensitivity 88% (95% CI 73-97%), positive predictive value 75% (95% CI 59-87%), and negative predictive value 98% (95% CI 94-99%).

conclusionUltrasound using a PUD is useful for evaluating dilatation of the collecting system, especially for ruling out its presence.

Indexed as

Point-of-Care SystemsAdultAgedAged, 80 and overFemaleHumansHydronephrosisKidneyMaleMiddle AgedProspective StudiesSensitivity and SpecificitySingle-Blind MethodUltrasonographyYoung AdultHydronephrosisPocket-sizedRenal collecting systemUltrasound