Evidence map›Paper›PMID 29703261›Full record

SynthesisBiomedical engineering online2018

Biomechanical analysis of plate systems for proximal humerus fractures: a systematic literature review.

Ali Jabran, Chris Peach, Lei Ren

Open access · goldAbstract readSystematic Review
In one paragraph

Synthesis in Biomedical engineering online, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 38 papers.

0numbers the graph read from it
0cells of the map it votes in
38citing papers in PubMed
6.6field-weighted citation impact, top 3% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

38 citing papers in PubMed, 76 citations in OpenAlex.

  1. Trial
  2. Article
  3. Review
  4. Article
  5. Review
  6. Article
  7. Article
  8. Proximal humerus fractures in the elderly: is there (still) a role for plate osteosynthesis?European journal of trauma and emergency surgery : official publication of the European Trauma Society · 2025
    Review
  9. Article
  10. Article
  11. Review
  12. Article
  13. Review
  14. Article
  15. Article
  16. Article
  17. Article
  18. Morphology and novel classification of proximal humeral fractures.Frontiers in bioengineering and biotechnology · 2024
    Article
  19. Article
  20. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors at 1 institution in 1 country.

Ali JabranSchool of Mechanical, Aerospace and Civil Engineering, University of Manchester, Manchester, UK.
Chris PeachSchool of Mechanical, Aerospace and Civil Engineering, University of Manchester, Manchester, UK.
Lei RenSchool of Mechanical, Aerospace and Civil Engineering, University of Manchester, Manchester, UK. lei.ren@manchester.ac.uk.
University of Manchester · GB

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundProximal humerus fractures are the third most common in the human body but their management remains controversial. Open reduction and internal fixation with plates is one of the leading modes of operative treatment for these fractures. The development of technologies and techniques for these plates, during the recent decades, promise a bright future for their clinical use. A comprehensive review of in vitro biomechanical studies is needed for the comparison of plates' mechanical performance and the testing methodologies. This will not only guide clinicians with plate selection but also with the design of future in vitro biomechanical studies. This review was aimed to systematically categorise and review the in vitro biomechanical studies of these plates based on their protocols and discuss their results. The technologies and techniques investigated in these studies were categorised and compared to reach a census where possible. METHODS AND

resultsWeb of Science and Scopus database search yielded 62 studies. Out of these, 51 performed axial loading, torsion, bending and/or combined bending and axial loading while 11 simulated complex glenohumeral movements by using tendons. Loading conditions and set-up, failure criteria and performance parameters, as well as results for each study, were reviewed. Only two studies tested four-part fracture model while the rest investigated two- and three-part fractures. In ten studies, synthetic humeri were tested instead of cadaveric ones. In addition to load-displacement data, three-dimensional motion analysis systems, digital image correlation and acoustic emission testing have been used for measurement.

conclusionsOverall, PHILOS was the most tested plate and locking plates demonstrated better mechanical performance than non-locking ones. Conflicting results have been published for their comparison with non-locking blade plates and polyaxial locking screws. Augmentation with cement [calcium phosphate or poly(methyl methacrylate)] or allografts (fibular and femoral head) was found to improve bone-plate constructs' mechanical performance. Controversy still lies over the use of rigid and semi-rigid implants and the insertion of inferomedial screws for calcar region support. This review will guide the design of in vitro and in silico biomechanical tests and also supplement the study of clinical literature.

Indexed as

Bone PlatesMechanical PhenomenaShoulder FracturesBiomechanical PhenomenaHumansBiomechanical testingLocking platesOpen reduction internal fixationProximal humerus fractures

Identifiers

PMID29703261
PMCPMC5923007
OpenAlexW2805747503

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.