Accuracy and precision of ScvO2 measured with the CeVOX-device: A prospective study in patients with a wide variation of ScvO2-values.

Alexander Herner, Bernhard Haller, Ulrich Mayr, Sebastian Rasch, Lea Offman, Roland Schmid, Wolfgang Huber

Abstract readRandomized Controlled TrialValidation Study
In one paragraph

Trial report in PloS one, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed.

  1. Do ScvOPhysiological reports · 2021
    Trial
  2. Article
  3. Article
  4. Article
  5. Review
  6. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Alexander HernerMedizinische Klinik und Poliklinik II, Klinikum rechts der Isar, Technische Universität München, München, Germany.
Bernhard HallerInstitut für Medizinische Statistik und Epidemiologie, Klinikum rechts der Isar, Technische Universität München, München, Germany.
Ulrich MayrMedizinische Klinik und Poliklinik II, Klinikum rechts der Isar, Technische Universität München, München, Germany.
Sebastian RaschMedizinische Klinik und Poliklinik II, Klinikum rechts der Isar, Technische Universität München, München, Germany.
Lea OffmanMedizinische Klinik und Poliklinik II, Klinikum rechts der Isar, Technische Universität München, München, Germany.
Roland SchmidMedizinische Klinik und Poliklinik II, Klinikum rechts der Isar, Technische Universität München, München, Germany.
Wolfgang HuberMedizinische Klinik und Poliklinik II, Klinikum rechts der Isar, Technische Universität München, München, Germany.ORCID 0000-0001-9086-7908

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

introductionCentral-venous oxygen saturation (ScvO2) is a key parameter of hemodynamic monitoring and has been suggested as therapeutic goal for resuscitation. Several devices offer continuous monitoring features. The CeVOX-device (Pulsion Medical Systems) uses a fibre-optic probe inserted through a conventional central-venous catheter (CVC) to obtain continuous ScvO2.

objectivesSince there is a lack of studies validating the CeVOX, we prospectively analyzed data from 24 patients with CeVOX-monitoring. To increase the yield of lower ScvO2-values, 12 patients were equipped with a femoral CVC.

methodsDuring the 8h study period ScvO2_CeVOX was documented immediately before withdrawal of blood to measure ScvO2 by blood gas analysis (ScvO2_BGA) 6min, 1h, 4h, 5h and 8h after the initial calibration. No further calibrations were performed.

resultsIn patients with jugular CVC (primary endpoint; 60 measurements), bias, lower and upper limits of agreement (LLOA; ULOA) and percentage error (PE) of the estimate of ScvO2 (ScvO2_CeVOX_jug) were acceptable with 0.45%, -13.0%, 13.9% and 16.6%, respectively. As supposed, ScvO2 was lower in the femoral compared to the jugular measurements (69.5±10.7 vs. 79.4±5.8%; p<0.001). While the bias (0.64%) was still acceptable, LLOA (-23.8%), ULOA (25.0%) and PE (34.5%) were substantially higher for femoral assessment of ScvO2 by the CeVOX (ScvO2_CeVOX_fem). Analysis of the entire data-pool with jugular as well as femoral CVCs allowed for a multivariate analysis which demonstrated that the position of the CVC per se was not independently associated with the bias ScvO2_CeVOX-ScvO2_BGA. The amount of the bias |ScvO2_CeVOX-ScvO2_BGA| was independently associated with the amount of the change of ScvO2_CeVOX compared to the initial calibration to ScvO2_BGA_baseline (|ScvO2_CeVOX-ScvO2_BGA_baseline|) as well as with low values of ScvO2_BGA_baseline. Furthermore, increasing time to the initial calibration was associated to the amount of the bias with borderline significance. A statistical model based on |ScvO2_CeVOX-ScvO2_BGA_baseline| and "time to last calibration" derived from an evaluation dataset (80 of 120 datasets, 16 of 24) provided a ROC-AUC of 0.903 to predict an amount of the bias |ScvO2_CeVOX-ScvO2_BGA| ≥5% in an independent validation group (40 datasets of 8 patients).

conclusionThese findings suggest that the CeVOX device is capable to detect stability or instability of ScvO2_BGA. ScvO2_CeVOX accurately estimates ScvO2_BGA in case of stable values. However, intermittent measurement of ScvO2_BGA and re-calibration should be performed in case of substantial changes in ScvO2_CeVOX compared to baseline. Therefore, continuous measurement of ScvO2 with the CeVOX cannot replace ScvO2_BGA in instable patients. On the other hand, CeVOX might be useful for the monitoring of stable patients as a pre-test tool for more differentiated monitoring in case of changes in ScvO2_CeVOX.

Indexed as

Central Venous CathetersAgedAged, 80 and overBlood Gas AnalysisFemaleFemoral VeinHumansJugular VeinsMaleMiddle AgedOxygenOxygen ConsumptionProspective StudiesRespiratory Distress SyndromeSepsisOxygen

Identifiers

PMID29664900
PMCPMC5903646