Evidence map›Paper›PMID 29480891›Full record

ArticleClinical orthopaedics and related research2018

High Revision Rate for Large-head Metal-on-metal THA at a Mean of 7.1 Years: A Registry Study.

Matti Seppänen, Inari Laaksonen, Pekka Pulkkinen, Antti Eskelinen, Ari-Pekka Puhto, Jukka Kettunen, Jarkko Leskinen, Mikko Manninen, Keijo Mäkelä

Open access · greenAbstract read
In one paragraph

Article in Clinical orthopaedics and related research, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 11 papers.

0numbers the graph read from it
0cells of the map it votes in
11citing papers in PubMed
2.8field-weighted citation impact, top 10% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

11 citing papers in PubMed, 19 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
  9. Article
  10. Article
  11. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors at 1 institution in 1 country.

Matti SeppänenM. Seppänen, I. Laaksonen, K. Mäkelä, Department of Orthopaedics and Traumatology, Turku University Hospital, Turku, Finland; and University of Turku, Turku, Finland P. Pulkkinen, Department of Public Health, Helsinki University, Helsinki, Finland A. Eskelinen, Department of Orthopaedics and Traumatology, Coxa Hospital for Joint Replacement, Tampere, Finland A.-P. Puhto, Department of Orthopaedics and Traumatology, Oulu University Hospital, Oulu, Finland J. Kettunen, Department of Orthopaedics and Traumatology, Kuopio University Hospital, Kuopio, Finland J. Leskinen, Department of Orthopaedics and Traumatology, Helsinki University Central Hospital, Helsinki, Finland M. Manninen, Orton Hospital, Helsinki, Finland.
Inari Laaksonen
Pekka Pulkkinen
Antti Eskelinen
Ari-Pekka Puhto
Jukka Kettunen
Jarkko Leskinen
Mikko Manninen
Keijo Mäkelä
Turku University Hospital · FI

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundLarge-diameter head metal-on-metal (MoM) THA has largely been abandoned as a result of higher than anticipated revision rates. However, the majority of these implants are still in situ. Although earlier reports from the Finnish Arthroplasty Register noted similar short-term survivorship between large-diameter head MoM THA and conventional cemented THA, longer term survivorship of this population is unclear. Although reported revision rates for this implant group have been high, the majority of these implants have not been revised and followup is important to improve long-term management. QUESTIONS/PURPOSES: The purposes of this study were (1) to compare the 10-year competing risk survivorship of large-diameter head MoM THA with the survivorship of conventional THA in the Finnish Arthroplasty Register; (2) to report the large-diameter head MoM THA survival at the manufacturer/brand level; and (3) to identify the most common reasons for revision of large-diameter head MoM THA in the Finnish Arthroplasty Register.

methodsThe six most commonly used large-diameter head (≥ 38 mm) MoM THA devices in Finland between years 2004 and 2013 were selected (n = 10,959 implants). The completeness of the Finnish Registry is > 95% in primary THA and patients are censored from the date of death or at the point of emigration; followup continued until the end of 2015. The conventional THA control group consisted of the two most frequently used devices (Vision/Bimetric and ABG II/ABG II) with metal-on-polyethylene or ceramic-on-ceramic bearing surfaces implanted between 2002 and 2013 (n = 5177). The study group was formed by selecting all pairs of large-diameter head MoM and reference THA protheses within the same age group ( < 49, 50-54, 55-59, 60-64, 65-69, 70-74, and 75+ years), sex, diagnosis (osteoarthritis, other), and hospital yearly operation count (< 100 operations yearly, ≥ 100 operations yearly); 5166 matched pairs were identified. Revision for any reason was considered as the failure endpoint of followup. Implant survival (the proportion not revised) was calculated from corresponding cumulative incidence function adjusted for patient death as a competing event for revision. Large-diameter head MoM implant group revision hazard ratios with 95% confidence intervals were estimated with age group, sex, diagnosis, and hospital yearly operation count as confounding factors in a Cox regression model.

resultsTen-year survivorship free from all-cause revision was lower for THAs that used a large-diameter femoral head than it was for the control group of conventional THA (83% [95% confidence interval {CI}, 82%-84%] versus 92% [95% CI, 91%-93%]). At the implant level, every large-diameter head MoM THA had a higher risk for revision compared with the conventional THA control group from the fourth postoperative year onward. The highest survival of MoM THA was 88% (95% CI, 86%-90%) for the ReCap/Bimetric and the lowest survival was 46% (95% CI, 41%-51%) for the recalled ASR with either the Summit® or Corail® stem. The most common revision reason in the MoM THA group was adverse reaction to metal debris, whereas dislocation was predominant in the conventional THA control group.

conclusionsThe revision rate for all large-diameter head MoM THAs in this timeframe in the Finnish Arthroplasty Register is unacceptably high and in our view supports the decision to abandon their use. In agreement with the directives of other national organizations, we recommend regular followup of all patients with large-diameter head MoM THA. Based on our results, strict guidelines for followup should be maintained over the lifetime of the implant to assess patient symptoms and recommend revision when indicated. LEVEL OF EVIDENCE: Level III, therapeutic study.

Indexed as

AdultAgedArthroplasty, Replacement, HipCeramicsFemaleFinlandHip ProsthesisHumansMaleMetal-on-Metal Joint ProsthesesMiddle AgedPolyethyleneProportional Hazards ModelsProsthesis DesignProsthesis FailureRegistriesPolyethylene

Identifiers

PMID29480891
PMCPMC6263568
OpenAlexW2809694668

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.