Evidence map›Paper›PMID 29476203›Full record

ArticleClinical research in cardiology : official journal of the German Cardiac Society2018

First experience with zero-fluoroscopic ablation for supraventricular tachycardias using a novel impedance and magnetic-field-based mapping system.

Katie A Walsh, Joseph Galvin, John Keaney, Edward Keelan, Gabor Szeplaki

Open access · hybridAbstract read
In one paragraph

Article in Clinical research in cardiology : official journal of the German Cardiac Society, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 27 papers, 2 of them syntheses that pooled it.

0numbers the graph read from it
0cells of the map it votes in
27citing papers in PubMed, 2 pooled it
2.7field-weighted citation impact, top 9% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

27 citing papers in PubMed, 2 syntheses or guidelines pooled it, 35 citations in OpenAlex.

  1. Pooled it
  2. Pooled it
  3. Randomized trial of intracardiac echocardiography-guided slow pathway ablation.Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2022
    Trial
  4. Review
  5. Article
  6. Article
  7. Article
  8. Article
  9. Article
  10. Article
  11. Article
  12. Article
  13. Article
  14. Review
  15. Article
  16. Article
  17. Observational
  18. Article
  19. Article
  20. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 2 institutions in 3 countries.

Katie A WalshHeart and Vascular Centre, Mater Private Hospital, 72 Eccles Street, Dublin 7, Ireland.
Joseph GalvinHeart and Vascular Centre, Mater Private Hospital, 72 Eccles Street, Dublin 7, Ireland.
John KeaneyHeart and Vascular Centre, Mater Private Hospital, 72 Eccles Street, Dublin 7, Ireland.
Edward KeelanHeart and Vascular Centre, Mater Private Hospital, 72 Eccles Street, Dublin 7, Ireland.
Gabor SzeplakiHeart and Vascular Centre, Mater Private Hospital, 72 Eccles Street, Dublin 7, Ireland. szeplaki.gabor@gmail.com.ORCID http://orcid.org/0000-0003-3795-1240
Mater Misericordiae University Hospital · IESemmelweis University · HU

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

aimsZero- and near-zero-fluoroscopic ablation techniques reduce the harmful effects of ionizing radiation during invasive electrophysiology procedures. We aimed to test the feasibility and safety of a zero-fluoroscopic strategy using a novel integrated magnetic and impedance-based electroanatomical mapping system for radiofrequency ablation (RFA) of supraventricular tachycardias (SVTs).

methodsWe retrospectively studied 92 consecutive patients undergoing electrophysiology studies with/without RFA for supraventricular tachycardia (SVT) performed by a single operator at a single center. The first 42 (Group 1) underwent a conventional fluoroscopic-guided approach and the second 50 (Group 2) underwent a zero-fluoroscopic approach using the Ensite Precision

resultsGroup 1 comprised 14 AV-nodal re-entrant tachycardia (AVNRT), 12 typical atrial flutter, 4 accessory pathway (AP), 2 atrial tachycardia (AT), and 9 diagnostic EP studies (EPS). Group 2 comprised 16 AVNRT, 17 atrial flutter, 6 AP, 3 AT, 2 AV-nodal ablations, and 7 EPS. A complete zero-fluoroscopic approach was achieved in 94% of Group 2 patients. All procedures were acutely successful, and no complications occurred. There was a significant reduction in fluoroscopy dose, dose area product, and time (p < 0.0001, for all), with no difference in procedure times. Ablation time for typical atrial flutter was shorter in Group 2 (p = 0.006).

conclusionsA zero-fluoroscopic strategy for diagnosis and treatment of SVTs using this novel 3D-electroanatomical mapping system is feasible in majority of patients, is safe, reduces ionizing radiation exposure, and does not compromise procedural times, success rates, or complication rates.

Indexed as

Electrophysiologic Techniques, CardiacAccessory Atrioventricular BundleAction PotentialsAdultAgedAtrial FlutterCatheter AblationFeasibility StudiesFemaleFluoroscopyHeart RateHumansMagneticsMaleMiddle AgedOperative TimeAblationENSITE precisionRadiationSupraventricular tachycardiaTypical atrial flutterZero-fluoroscopy

Identifiers

PMID29476203
PMCPMC6002461
OpenAlexW2792627404

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.