ArticleKnee surgery, sports traumatology, arthroscopy : official journal of the ESSKA2018
Significant differences between manufacturer and surgeon in the accuracy of final component size prediction with CT-based patient-specific instrumentation for total knee arthroplasty.
Article in Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 10 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
10 citing papers in PubMed, 1 synthesis or guideline pooled it, 21 citations in OpenAlex.
- 3D imaging-based AI models outperform demographic models and excel in tibial sizing compared with 2D models in total knee arthroplasty planning: A systematic review.Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA · 2026Pooled it
- Patient-specific Instrumentation Affects Rotational Alignment of the Femoral Component in Total Knee Arthroplasty: A Prospective Randomized Controlled Trial.Orthopaedic surgery · 2019Trial
- Comparison between robotic-assisted and navigation-assisted total knee arthroplasty shows comparable outcomes: A systematic review and meta-analysis.Journal of orthopaedics · 2025Review
- The Accuracy of CT-Based Three-Dimensional Templating in Predicting Implant Sizes in Patients Undergoing Robot-Assisted Total Knee Arthroplasty.Indian journal of orthopaedics · 2024Article
- Reliable prediction of implant size and axial alignment in AI-based 3D preoperative planning for total knee arthroplasty.Scientific reports · 2024Article
- Predictability of implant sizes during cruciate-retaining total knee arthroplasty using an image-free hand-held robotic system.Journal of robotic surgery · 2024Article
- Navigation and robotics improved alignment compared with PSI and conventional instrument, while clinical outcomes were similar in TKA: a network meta-analysis.Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA · 2022Review
- Artificial Intelligence Based Patient-Specific Preoperative Planning Algorithm for Total Knee Arthroplasty.Frontiers in robotics and AI · 2022Article
- High intra- and inter-observer reliability of planning implant size in MRI-based patient-specific instrumentation for total knee arthroplasty.Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA · 2021Article
- What future in the treatment of osteochondral knee defects?Annals of translational medicine · 2018Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors at 2 institutions in 2 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposePatient-specific instrumentation (PSI) for total knee arthroplasty (TKA) may improve component sizing. Little has been reported about accuracy of the default plan created by the manufacturer, especially for CT-based PSI. The goal of this study was to evaluate the reliability of this plan and the impact of the surgeon's changes on the final accuracy of the guide sizes.
methodsForty-five patients eligible for primary TKA were prospectively enrolled. The planned implant sizes were prospectively recorded from the initial manufacturer's proposal and from the final plan adjusted in light of the surgeon's evaluation; these two sizes where then compared to the actually implanted sizes. Fisher's exact test was used to test differences for categorical variables. Agreement between pre-operative plans and final implant was evaluated with the Bland-Altman method.
resultsThe manufacturer's proposal differed from the final implant in 9 (20.0%) femoral and 23 (51.1%) tibial components, while the surgeon's plan in 6 (13.3%, femoral) and 12 (26.7%, tibial). Modifications in the pre-operative plan were carried out for five (11.1%) femoral and 23 (51.1%) tibial components (p = 0.03). Appropriate modification occurred in 22 (88.0%) and 19 (76.0%) cases of femoral and tibial changes. The agreement between the manufacturer's and the surgeon's pre-operative plans was poor, especially with regard to tibial components.
conclusionThe surgeon's accuracy in predicting the final component size was significantly different from that of the manufacturer and changes in the initial manufacturer's plan were necessary to get an accurate pre-operative plan of the implant sizes. CLINICAL RELEVANCE: Careful evaluation of the initial manufacturer's plan by an experienced knee surgeon is mandatory when planning TKA with CT-based PSI. LEVEL OF EVIDENCE: II.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials