Are there increased periprocedural complications with the MRI-conditional Medtronic Revo SureScan Pacing System? : A meta-analysis.

M Shurrab, A Kaoutskaia, A Baranchuk, C Lau, T Singarajah, I Lashevsky, D Newman, J S Healey, E Crystal

Abstract readReview
In one paragraph

Review in Netherlands heart journal : monthly journal of the Netherlands Society of Cardiology and the Netherlands Heart Foundation, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed.

  1. Article
  2. Review
  3. Magnetic resonance imaging in patients with cardiac implantable electronic devices.Netherlands heart journal : monthly journal of the Netherlands Society of Cardiology and the Netherlands Heart Foundation · 2018
    Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors.

M ShurrabArrhythmia Services, Schulich Heart Centre, Sunnybrook Health Sciences Centre, University of Toronto, Toronto, Ontario, Canada. shurrabm@hotmail.com.
A KaoutskaiaArrhythmia Services, Schulich Heart Centre, Sunnybrook Health Sciences Centre, University of Toronto, Toronto, Ontario, Canada.
A BaranchukDivision of Cardiology, Kingston General Hospital, Queen's University, Kingston, Ontario, Canada.
C LauArrhythmia Services, Schulich Heart Centre, Sunnybrook Health Sciences Centre, University of Toronto, Toronto, Ontario, Canada.
T SingarajahArrhythmia Services, Schulich Heart Centre, Sunnybrook Health Sciences Centre, University of Toronto, Toronto, Ontario, Canada.
I LashevskyArrhythmia Services, Schulich Heart Centre, Sunnybrook Health Sciences Centre, University of Toronto, Toronto, Ontario, Canada.
D NewmanArrhythmia Services, Schulich Heart Centre, Sunnybrook Health Sciences Centre, University of Toronto, Toronto, Ontario, Canada.
J S HealeyPopulation Health Research Institute, McMaster University, Hamilton, Ontario, Canada.
E CrystalArrhythmia Services, Schulich Heart Centre, Sunnybrook Health Sciences Centre, University of Toronto, Toronto, Ontario, Canada.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundThe use of magnetic resonance imaging (MRI)-conditional permanent pacemakers has increased significantly. In this meta-analysis, we examine the safety of MRI-conditional pacing systems in comparison with conventional systems.

methodsAn electronic search was performed using major databases, including studies that compared the outcomes of interest between patients receiving MRI-conditional pacemakers (MRI group) versus conventional pacemakers (control group).

resultsSix studies (5 retrospective and 1 prospective non-randomised) involving 2,118 adult patients were identified. The MRI-conditional pacemakers, deployed in 969 patients, were all from a single manufacturer (Medtronic Pacing System with 5086 leads). The rate of pacemaker lead dislodgement (atrial and ventricular) was significantly higher in the MRI group (3% vs. 1%, OR 2.47 (95% CI 1.26; 4.83), p = 0.008). The MRI group had a significantly higher rate of pericardial complications (2% vs. 1%, OR 4.23 (95% CI 1.18; 15.10), p = 0.03) and a numerically higher overall complication rate in comparison with the conventional group (6% vs. 3%, OR 2.02 (95% CI 0.88; 4.66), p = 0.10) but this was not statistically significant.

conclusionsIn this meta-analysis, the rates of pacemaker lead dislodgement and pericardial complications were significantly higher with the Medtronic MRI-conditional pacing system.