Polyethylene Wear Increases in Liners Articulating With Scratched Oxidized Zirconium Femoral Heads.
Article in Clinical orthopaedics and related research, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
8 citing papers in PubMed, 16 citations in OpenAlex.
- Dramatic Failure of an OXINIUM Total Knee Arthroplasty With a Massive Pseudotumor Formation.Arthroplasty today · 2024Article
- Third body damage and wear in arthroplasty bearing materials: A review of laboratory methods.Biomaterials and biosystems · 2021Review
- The Oxinium Arthrogram: A Sign of Oxidized Zirconium Implant Failure.Arthroplasty today · 2021Article
- Oxidized Zirconium Total Knee Arthroplasty Implant Failure in a Patient With Knee Instability.Arthroplasty today · 2020Article
- Protrusion of a ceramic femoral head through the acetabular metallic cup in total-hip arthroplasty: A case report.Medicine · 2020Article
- Long-term performance of oxidized zirconium on conventional and highly cross-linked polyethylene in total hip arthroplasty.SICOT-J · 2020Article
- MRI of THA Correlates With Implant Wear and Tissue Reactions: A Cross-sectional Study.Clinical orthopaedics and related research · 2019Article
- Editor's Spotlight/Take 5: Polyethylene Wear Increases in Liners Articulating With Scratched Oxidized Zirconium Femoral Heads.Clinical orthopaedics and related research · 2018Article
Corrections and comments
- Commented on by
Authors and funding
5 authors at 1 institution in 2 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundOxidized zirconium (OxZr) femoral heads were introduced in hip arthroplasty to reduce wear of the polyethylene compared with metallic heads and to reduce fracture risk compared with ceramic heads. Severe scratches have been reported on OxZr heads in patients undergoing revision for instability, but whether these scratches contribute to increased acetabular polyethylene wear remains unclear. QUESTIONS/PURPOSES: (1) How is the polyethylene of the acetabular liner affected by damage on the opposing OxZr head? (2) How does damage to the head affect the degree of polyethylene wear?
methodsWe assessed damage and deformation on all retrieved highly crosslinked liners that had articulated against OxZr heads collected at one institution between 2006 and 2013. Two observers used a visual subjective scoring system to assess polyethylene damage on the surface of the 42 retrieved liners. Polyethylene components were also laser scanned to measure dimensional changes to the liner. These outcomes were compared with the severity of scratching on the surface of the articulating OxZr head. We also used a 12-station hip simulator to measure wear over five million cycles (MCs) of pristine liners articulating against nine retrieved OxZr heads with varying degrees of scratching representing a spectrum of little to severe damage and three retrieved ceramic heads with severe metal transfer.
resultsSeventeen of the OxZr heads showed severe damage, of which 14 heads had been revised for dislocation. The retrieved liners that had articulated with these heads had greater damage scores for abrasion (mean score 0.4 versus 1.6; p = 0.008) and embedded debris (mean score 0.4 versus 1.4; p = 0.006) compared with liners that had articulated with less damaged heads. Four severely damaged OxZr heads wore at a higher rate than the others in the study with weight loss of 37.7 mg, 30.0 mg, 14.4 mg, and 2.6 mg after the first MC and a steady increase through testing to 5 MCs. Conversely, neither OxZr heads with less damage nor ceramic heads with severe metal transfer produced appreciable wear.
conclusionsSurface scratching of OxZr heads from recurrent dislocation and reduction maneuvers leads to increased wear of the crosslinked polyethylene used as a bearing surface. CLINICAL RELEVANCE: Patients in whom such instability occurs in the presence of an OxZr head should be followed closely for the possibility of more severe wear.