Evidence map›Paper›PMID 29318353›Full record

Observational studyClinical neuroradiology2019

The Barrel Vascular Reconstruction Device : A Retrospective, Observational Multicentric Study.

C Kabbasch, A Mpotsaris, V Maus, J C Altenbernd, C Loehr

Abstract readMulticenter StudyObservational Study
PubMed Publisher
In one paragraph

Observational study in Clinical neuroradiology, 2019. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 10 papers.

0numbers the graph read from it
0cells of the map it votes in
10citing papers in PubMed
2.0field-weighted citation impact, top 14% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

10 citing papers in PubMed, 16 citations in OpenAlex.

  1. Article
  2. Review
  3. Review
  4. Review
  5. Review
  6. Article
  7. Analysis of branch artery orifice angulation: Feasibility of the shelf technique for the treatment of wide-neck bifurcation aneurysms.Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences · 2021
    Article
  8. Review
  9. Neuroform Atlas stent deployment through the Barrel vascular reconstruction device for the treatment of a wide-necked acutely ruptured basilar aneurysm.Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences · 2019
    Article
  10. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 3 institutions in 1 country.

C KabbaschDepartment of Radiology and Neuroradiology, University Hospital of Cologne, Kerpener Straße 62, 50924, Cologne, Germany. christoph.kabbasch@uk-koeln.de.ORCID http://orcid.org/0000-0003-3712-2258
A MpotsarisDepartment of Radiology and Neuroradiology, University Hospital of Cologne, Kerpener Straße 62, 50924, Cologne, Germany.
V MausDepartment of Radiology and Neuroradiology, University Hospital of Cologne, Kerpener Straße 62, 50924, Cologne, Germany.
J C AltenberndDepartment of Radiology and Neuroradiology, Klinikum Vest - Knappschaftskrankenhaus Recklinghausen, 45657, Recklinghausen, Germany.
C LoehrDepartment of Radiology and Neuroradiology, Klinikum Vest - Knappschaftskrankenhaus Recklinghausen, 45657, Recklinghausen, Germany.
Klinikum Vest · DEUniversity Hospital Cologne · DEUniversitätsklinikum Aachen · DE

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

background and purposeThe Barrel device is an electrolytically detachable laser cut, closed-cell microstent that is used for neck reconstruction in wide-necked bifurcation and branching aneurysms to support coiling. The key feature is a barrel section that herniates over the aneurysmal ostium. The objective was to evaluate the safety, feasibility and the immediate and mid-term occlusion results of this new device. MATERIAL AND

methodsThe databases of two tertiary care centers were screened for all Barrel-based stent-assisted intracranial coil embolization of wide-necked aneurysms between June 2015 and September 2016. Case files and imaging data were retrospectively analyzed for angiographic and clinical outcome parameters, including immediate and mid-term modified Raymond-Roy aneurysm occlusion classification (RROC) rates and procedural complications.

resultsA total of 21 patients comprising 21 intracranial aneurysms (20 unruptured, 1 ruptured) were treated with the Barrel device and additional coiling of the aneurysm sac. All aneurysms were wide-necked, saccular bifurcation aneurysms defined by a dome/neck ratio ≤2. Immediate complete occlusion (RROC1) was observed in 19/21 (90%). An intra-interventional in-stent thrombus formation in two cases (10%) was medically resolved without neurological sequelae. A single case of symptomatic in-stent stenosis (5%) was cleared by balloon angioplasty. Follow-up (FU) was available in 20/21 cases (95%) after a median of 282 days (range: 17-591 days). At follow-up 19/20 aneurysms (95%) were completely occluded (RROC1).

conclusionThe Barrel device showed a satisfactory safety profile and a promising rate of immediate and mid-term complete aneurysm occlusion for stent-assisted coil embolization of wide-necked intracranial bifurcation aneurysms, warranting further investigation of the device.

Indexed as

StentsAdultAgedAged, 80 and overAngiography, Digital SubtractionEmbolization, TherapeuticEndovascular ProceduresFeasibility StudiesFemaleFollow-Up StudiesHumansIntracranial AneurysmMagnetic Resonance AngiographyMaleMiddle AgedModels, BiologicalPlatelet Aggregation InhibitorsBifurcation stentCoilingDeviceEndovascularWide-necked aneurysm

Identifiers

PMID29318353
OpenAlexW2783003802

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.