ArticleNuclear medicine and molecular imaging2017
Evaluation of Cadmium-Zinc-Telluride Detector-based Single-Photon Emission Computed Tomography for Nuclear Cardiology: a Comparison with Conventional Anger Single-Photon Emission Computed Tomography.
Article in Nuclear medicine and molecular imaging, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 14 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
14 citing papers in PubMed.
- QuantitativeEJNMMI physics · 2026Article
- Single photon γ-ray imaging with high energy and spatial resolution perovskite semiconductor for nuclear medicine.Nature communications · 2025Article
- Accuracy of a whole-body single-photon emission computed tomography with a thallium-bromide detector: Verification via Monte Carlo simulations.Medical physics · 2025Article
- Recent advances in radionuclide medical imaging techniques.Frontiers in medicine · 2025Review
- True Opinions of Nuclear Cardiology Practitioners on Semiconductor SPECT Systems: A Multi-Center Survey-Based Review.Annals of nuclear cardiology · 2024Review
- Comparison of positional artifacts in myocardial perfusion imaging in supine and semi-reclining position using dedicated D-SPECT cardiac camera: validation using CT based attenuation correction.Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology · 2023Article
- Head-to-Head Comparison of CZT-SPECT and SPECT/CT Myocardial Perfusion Imaging: Interobserver and Intraobserver Agreement and Diagnostic Performance.Life (Basel, Switzerland) · 2023Article
- Review
- Simultaneous dual-tracerJournal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology · 2021Article
- Experimental evaluation of the GE NM/CT 870 CZT clinical SPECT system equipped with WEHR and MEHRS collimator.Journal of applied clinical medical physics · 2021Article
- High-quality brain perfusion SPECT images may be achieved with a high-speed recording using 360° CZT camera.EJNMMI physics · 2020Article
- [Advancement of imaging technology for coronary microcirculation dysfunction assessment].Sheng wu yi xue gong cheng xue za zhi = Journal of biomedical engineering = Shengwu yixue gongchengxue zazhi · 2020Review
- Optimal composition and position of histidine-containing tags improves biodistribution ofScientific reports · 2019Article
- Comparative Cardiac Phantom Study Using Tc-99m/I-123 and Tl-201/I-123 Tracers with Cadmium-Zinc-Telluride Detector-Based Single-Photon Emission Computed Tomography.Nuclear medicine and molecular imaging · 2019Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeThe differences in performance between the cadmium-zinc-telluride (CZT) camera or collimation systems and conventional Anger single-photon emission computed tomography (A-SPECT) remain insufficient from the viewpoint of the user. We evaluated the performance of the D-SPECT (Spectrum Dynamics, Israel) system to provide more information to the cardiologist or radiological technologist about its use in the clinical field. MATERIALS AND
methodsThis study evaluated the performance of the D-SPECT system in terms of energy resolution, detector sensitivity, spatial resolution, modulation transfer function (MTF), and collimator resolution in comparison with that of A-SPECT (Bright-View, Philips, Japan). Energy resolution and detector sensitivity were measured for Tc-99m, I-123, and Tl-201. The SPECT images produced by both systems were evaluated visually using the anthropomorphic torso phantom.
resultsThe energy resolution of D-SPECT with Tc-99m and I-123 was approximately two times higher than that of A-SPECT. The detector sensitivity of D-SPECT was higher than that of A-SPECT (Tc-99m: 4.2 times, I-123: 2.2 times, and Tl-201: 5.9 times). The mean spatial resolution of D-SPECT was two times higher than that of A-SPECT. The MTF of D-SPECT was superior to that of the A-SPECT system for all frequencies. The collimator resolution of D-SPECT was lower than that of A-SPECT; however, the D-SPECT images clearly indicated better spatial resolution than the A-SPECT images.
conclusionThe energy resolution, detector sensitivity, spatial resolution, and MTF of D-SPECT were superior to those of A-SPECT. Although the collimator resolution was lower than that of A-SPECT, the D-SPECT images were clearly of better quality.