Evidence map›Paper›PMID 29242727›Full record

ArticleNuclear medicine and molecular imaging2017

Evaluation of Cadmium-Zinc-Telluride Detector-based Single-Photon Emission Computed Tomography for Nuclear Cardiology: a Comparison with Conventional Anger Single-Photon Emission Computed Tomography.

Takanaga Niimi, Mamoru Nanasato, Mitsuo Sugimoto, Hisatoshi Maeda

Abstract read
In one paragraph

Article in Nuclear medicine and molecular imaging, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 14 papers.

0numbers the graph read from it
0cells of the map it votes in
14citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

14 citing papers in PubMed.

  1. QuantitativeEJNMMI physics · 2026
    Article
  2. Article
  3. Article
  4. Review
  5. Review
  6. Article
  7. Article
  8. Review
  9. Simultaneous dual-tracerJournal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology · 2021
    Article
  10. Article
  11. Article
  12. [Advancement of imaging technology for coronary microcirculation dysfunction assessment].Sheng wu yi xue gong cheng xue za zhi = Journal of biomedical engineering = Shengwu yixue gongchengxue zazhi · 2020
    Review
  13. Article
  14. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Takanaga NiimiDepartment of Radiological Technology, Nagoya Daini Red Cross Hospital, Takanaga Niimi, 2-9 Myouken-cho, Showa-ku, Nagoya, 466-8650 Japan.ORCID 0000-0002-4098-9402
Mamoru NanasatoCardiovascular Center, Nagoya Daini Red Cross Hospital, 2-9 Myouken-cho, Showa-ku, Nagoya, 466-8650 Japan.
Mitsuo SugimotoDepartment of Radiological Technology, Nagoya Daini Red Cross Hospital, Takanaga Niimi, 2-9 Myouken-cho, Showa-ku, Nagoya, 466-8650 Japan.
Hisatoshi MaedaDepartment of Radiological Technology, Nagoya University School of Health Sciences, 1-1-20 Daiko-minami, Higashi-ku, Nagoya, 461-8673 Japan.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeThe differences in performance between the cadmium-zinc-telluride (CZT) camera or collimation systems and conventional Anger single-photon emission computed tomography (A-SPECT) remain insufficient from the viewpoint of the user. We evaluated the performance of the D-SPECT (Spectrum Dynamics, Israel) system to provide more information to the cardiologist or radiological technologist about its use in the clinical field. MATERIALS AND

methodsThis study evaluated the performance of the D-SPECT system in terms of energy resolution, detector sensitivity, spatial resolution, modulation transfer function (MTF), and collimator resolution in comparison with that of A-SPECT (Bright-View, Philips, Japan). Energy resolution and detector sensitivity were measured for Tc-99m, I-123, and Tl-201. The SPECT images produced by both systems were evaluated visually using the anthropomorphic torso phantom.

resultsThe energy resolution of D-SPECT with Tc-99m and I-123 was approximately two times higher than that of A-SPECT. The detector sensitivity of D-SPECT was higher than that of A-SPECT (Tc-99m: 4.2 times, I-123: 2.2 times, and Tl-201: 5.9 times). The mean spatial resolution of D-SPECT was two times higher than that of A-SPECT. The MTF of D-SPECT was superior to that of the A-SPECT system for all frequencies. The collimator resolution of D-SPECT was lower than that of A-SPECT; however, the D-SPECT images clearly indicated better spatial resolution than the A-SPECT images.

conclusionThe energy resolution, detector sensitivity, spatial resolution, and MTF of D-SPECT were superior to those of A-SPECT. Although the collimator resolution was lower than that of A-SPECT, the D-SPECT images were clearly of better quality.

Indexed as

Anger SPECTCollimation systemCZT cameraD-SPECTNuclear cardiology

Identifiers

PMID29242727
PMCPMC5721085

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.