ArticleEuropean radiology2018
Evaluation of an adaptive detector collimation for prospectively ECG-triggered coronary CT angiography with third-generation dual-source CT.
Article in European radiology, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
6 citing papers in PubMed, 13 citations in OpenAlex.
- Effect of different breath-holding methods on scanning length and radiation dose of 320-row volumetric coronary CT angiography.Japanese journal of radiology · 2025Trial
- Reduction of cardiac motion artifact in step-and-shoot coronary CT angiography with third-generation as compared with second-generation dual-source CT scanners.Diagnostic and interventional radiology (Ankara, Turkey) · 2021Trial
- Triple-rule-out CT angiography using two axial scans with 16 cm wide-detector for radiation dose reduction.European radiology · 2018Trial
- Contrast medium administration with a body surface area protocol in step-and-shoot coronary computed tomography angiography with dual-source scanners.Scientific reports · 2020Article
- Third-generation dual-source dual-energy CT in pediatric congenital heart disease patients: state-of-the-art.La Radiologia medica · 2019Review
- Reducing both radiation and contrast doses in coronary CT angiography in lean patients on a 16-cm wide-detector CT using 70 kVp and ASiR-V algorithm, in comparison with the conventional 100-kVp protocol.European radiology · 2019Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors at 5 institutions in 3 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectivesTo investigate the impact of an adaptive detector collimation on the dose parameters and accurateness of scan length adaption at prospectively ECG-triggered sequential cardiac CT with a wide-detector third-generation dual-source CT.
methodsIdeal scan lengths for human hearts were retrospectively derived from 103 triple-rule-out examinations. These measures were entered into the new scanner operated in prospectively ECG-triggered sequential cardiac scan mode with three different detector settings: (1) adaptive collimation, (2) fixed 64 × 0.6-mm collimation, and (3) fixed 96 × 0.6-mm collimation. Differences in effective scan length and deviation from the ideal scan length and dose parameters (CTDIvol, DLP) were documented.
resultsThe ideal cardiac scan length could be matched by the adaptive collimation in every case while the mean scanned length was longer by 15.4% with the 64 × 0.6 mm and by 27.2% with the fixed 96 × 0.6-mm collimation. While the DLP was almost identical between the adaptive and the 64 × 0.6-mm collimation (83 vs. 89 mGycm at 120 kV), it was 62.7% higher with the 96 × 0.6-mm collimation (135 mGycm), p < 0.001.
conclusionThe adaptive detector collimation for prospectively ECG-triggered sequential acquisition allows for adjusting the scan length as accurate as this can only be achieved with a spiral acquisition. This technique allows keeping patient exposure low where patient dose would significantly increase with the traditional step-and-shoot mode. KEY POINTS: • Adaptive detector collimation allows keeping patient exposure low in cardiac CT. • With novel detectors the desired scan length can be accurately matched. • Differences in detector settings may cause 62.7% of excessive dose.