Evidence map›Paper›PMID 29216858›Full record

ArticleBMC surgery2017

Use of a hydrophilic coating wire reduces significantly the rate of central vein punctures and the incidence of pneumothorax in totally implantable access port (TIAP) surgery.

Georgios Polychronidis, Roland Hennes, Cosima Engerer, Phillip Knebel, Daniel Schultze, Thomas Bruckner, Beat P Müller-Stich, Lars Fischer

Open access · goldAbstract read
In one paragraph

Article in BMC surgery, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
2.0field-weighted citation impact, top 13% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed, 9 citations in OpenAlex.

  1. Outcomes of Intravenous Normal Saline Infusion Pre-Cardiac Implantable Electronic Devices Versus No Infusion in Fasting.Annals of noninvasive electrocardiology : the official journal of the International Society for Holter and Noninvasive Electrocardiology, Inc · 2026
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 3 institutions in 1 country.

Georgios Polychronidis
Roland Hennes
Cosima EngererDepartment of Surgery, University of Heidelberg, INF 110, 69120, Heidelberg, Germany.
Phillip KnebelDepartment of Surgery, University of Heidelberg, INF 110, 69120, Heidelberg, Germany.
Daniel SchultzeDepartment of Surgery, University of Heidelberg, INF 110, 69120, Heidelberg, Germany.
Thomas BrucknerDepartment of Medical Biometry, Institute of Medical Biometry and Informatics (IMBI), Im Neuenheimer Feld 130.3, 69120, Heidelberg, Germany.
Beat P Müller-StichDepartment of Surgery, University of Heidelberg, INF 110, 69120, Heidelberg, Germany.
Lars FischerDepartment of Surgery, University of Heidelberg, INF 110, 69120, Heidelberg, Germany. L.Fischer@klinikum-mittelbaden.de.
Heidelberg University · DEKlinikum Mittelbaden · DEZimmer Biomet (Germany) · DE

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundInsertion of a Totally Implantable Access Port (TIAP) can be performed either via Central Vein Puncture (CVP) or Brachiocephalic Vein Cut-down (venous section-VS). The primary success rate of TIAP implantation using VS rarely ever achieves 100%. The objective of this study was to describe a modified VS approach using a hydrophilic coated wire (TVS).

methodsFrom 01.01.2015 to 31.12.2015, all patients receiving TIAP implantations were screened. During this time, all patients in whom the primary VS procedure was found to be unsuccessful were analysed.

resultsIn 2015, 1152 patients had TIAP implantations performed by 24 different surgeons. Of these, 277 patients needed a second line rescue strategy either by CVP (n = 69) or TVS (n = 208). There were no statistically significant differences regarding demographics or indication for TIAP implantation between CVP and TVS. The operation time and the qualification of the operating surgeon between CVP and TVS did not differ significantly. After the introduction of the guidewire with a hydrophilic coated wire, the need for CVP decreased significantly from 12.7% to 8.8% (p < 0.0001). In patients receiving CVP as a second line rescue strategy, the incidence of pneumothorax (n = 3 patients (4.3%)) was significantly higher compared to patients with TVS as a second line rescue strategy (n = 1 patient (0.48%), p = 0.02).

conclusionThe use of a hydrophilic coated wire significantly decreased the number of CVP and the incidence of pneumothorax. TVS is a safe and successful second-line rescue strategy.

Indexed as

Central Venous CathetersInfusion Pumps, ImplantableCatheterization, Central VenousEquipment DesignFemaleHumansMaleMiddle AgedPneumothoraxProstheses and ImplantsVeinsWounds, StabGuide wirePneumothoraxTotally implantable access portVenous cut-down

Identifiers

PMID29216858
PMCPMC5721482
OpenAlexW2772538665

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.