Combined effects of leaks, respiratory system properties and upper airway patency on the performance of home ventilators: a bench study.

Kaixian Zhu, Claudio Rabec, Jésus Gonzalez-Bermejo, Sébastien Hardy, Sami Aouf, Pierre Escourrou, Gabriel Roisman

Open access · goldAbstract readComparative Study
In one paragraph

Article in BMC pulmonary medicine, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 12 papers.

0numbers the graph read from it
0cells of the map it votes in
12citing papers in PubMed
1.0field-weighted citation impact, top 21% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

12 citing papers in PubMed, 36 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Are all ventilators for NIV performing the same? A bench analysis.Journal of clinical monitoring and computing · 2023
    Article
  5. Review
  6. Ventilators, Settings, Autotitration Algorithms.Journal of clinical medicine · 2023
    Review
  7. Monitoring Systems in Home Ventilation.Journal of clinical medicine · 2023
    Review
  8. Article
  9. Article
  10. Article
  11. Review
  12. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 5 institutions in 2 countries.

Kaixian ZhuCentre Explor, Air Liquide Healthcare, 28 rue d'Arcueil, 94250, Gentilly, France. kaixian.zhu@gmail.com.ORCID http://orcid.org/0000-0001-8599-9228
Claudio RabecService de Pneumologie et Soins Intensifs Respiratoires, Centre Hospitalier Universitaire Dijon Bourgogne, 14 rue Paul Gaffarel, F-21079, Dijon, France.
Jésus Gonzalez-BermejoSorbonne Universités, UPMC Univ Paris 6, INSERM, UMRS1158 Neurophysiologie Respiratoire Expérimentale et Clinique, Paris, France.
Sébastien HardyCentre Explor, Air Liquide Healthcare, 28 rue d'Arcueil, 94250, Gentilly, France.
Sami AoufCentre Explor, Air Liquide Healthcare, 28 rue d'Arcueil, 94250, Gentilly, France.
Pierre EscourrouService des Explorations Fonctionnelles Multidisciplinaires, AP-HP, Hôpital Antoine-Béclère, 157 rue de la Porte de Trivaux, 92140, Clamart, France.
Gabriel RoismanService des Explorations Fonctionnelles Multidisciplinaires, AP-HP, Hôpital Antoine-Béclère, 157 rue de la Porte de Trivaux, 92140, Clamart, France.
Air Liquide (France) · FRAssistance Publique – Hôpitaux de Paris · FRAir Liquide (United Kingdom) · GBCHU Dijon Bourgogne · FRInserm · FR

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundCombined effects of leaks, mechanical property of respiratory system and upper airway (UA) patency on patient-ventilator synchrony (PVA) and the level of clinically "tolerable" leaks are not well established in home ventilators.

methodsWe comparatively assessed on a bench model, the highest leak level tolerated without inducing significant asynchrony ("critical leak") in three home ventilators (Astral 150, Trilogy 100 and Vivo 60; noted as A150, T100 and V60 respectively) subjected to three simulated diseased respiratory conditions: chronic obstructive pulmonary disease (COPD), obesity hypoventilation (OHS) and neuromuscular disorders (NMD), with both open and closed UA. Also, total leak values in the device reports were compared to the bench-measured values.

resultsWith open UA, all ventilators were able to avoid asynchrony up to a 30 L/min leak and even to 55 L/min in some cases. UA closure and respiratory diseases especially OHS influenced PVA. With closed UA, the critical leak of A150 and T100 remained higher than 55 L/min in COPD and OHS, while for V60 decreased to 41 and 33 L/min respectively. In NMD with closed UA, only T100 reached a high critical leak of 69 L/min. Besides, inspiratory trigger sensitivity change was often necessary to avoid PVA.

conclusionsHome ventilators were able to avoid PVA in high-level leak conditions. However, asynchrony appeared in cases of abnormal mechanical properties of respiratory system or closed UA. In case of closed UA, the EPAP should be adjusted prior to the inspiratory trigger.

trial registrationNot applicable.

Indexed as

Ventilators, MechanicalEquipment DesignEquipment Failure AnalysisHumansModels, BiologicalNeuromuscular DiseasesNoninvasive VentilationObesity Hypoventilation SyndromePulmonary Disease, Chronic ObstructiveRespiratory MechanicsSelf CareTracheaNon-invasive ventilationPatient-ventilator asynchronyUnintentional leakUpper airway patency

Identifiers

PMID29157220
PMCPMC5697337
OpenAlexW2768410550

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.