[Analysis of the accuracy of fit of five anatomically precontoured clavicle plate systems].
Article in Der Unfallchirurg, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- The first 100 patients treated with a new anatomical pre-contoured locking plate for clavicular midshaft fractures.BMC musculoskeletal disorders · 2019Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundA common method in surgical treatment of midshaft clavicular fractures is the clavicle plating system. In addition to traditional osteosynthetic devices, anatomically preformed plate systems also exist. Advantages are described in greater accuracy of fit and less irritation of surrounding soft tissues. The purpose of this study was to compare the anatomical fit of five different precontoured clavicle plating systems. MATERIAL AND
methodsThe anatomical fit of five different types of anatomical precontoured clavicle plates, 3 times VariaxSystem® (Stryker, Kalamazoo, MI), Meves® Plate (Ulrich medical Ulm, Germany) and LCP clavicle plate (Synthes, Bettlach, Switzerland) were investigated in 20 embalmed human cadaveric clavicles. An imprint of the space between the well-positioned plate and the clavicle was obtained using a silicone mass and the silicone imprint was digitally measured. Additionally, the anatomical fit was evaluated by three investigators following a standardized protocol (+2 to -2 points at lateral, midshaft and medial clavicle portions).
resultsThe first three ranks went to the Stryker Variax-plates. They showed the least distance (Stryker 628027: 7‑hole 1.44 mm, low curvature 0.93 mm-2.36 mm, 2. Stryker 628028: 8‑hole low curvature 1.68 mm, 1.03 mm-2.4 mm and 3. Stryker 628128: 8‑hole high curvature 1.87 mm, range 1.09 mm-3.1 mm) The evaluation of the anatomical fit by the investigators was in agreement with the measurement results.
conclusionAlthough there was no complete congruency between the plates and the clavicle, all clavicle plates investigated in this study presented a reasonable anatomical shape. The 7‑hole VariAx Stryker plate with slight curvature showed the best anatomical fit. A low profile and optimized anatomical precontouring can minimize irritation of the surrounding soft tissues and should be considered in plate design and implant choice.