Evidence map›Paper›PMID 29048330›Full record

ArticlePhysics in medicine and biology2017

Calibration and analysis of a multimodal micro-CT and structured light imaging system for the evaluation of excised breast tissue.

David M McClatchy, Elizabeth J Rizzo, Jeff Meganck, Josh Kempner, Jared Vicory, Wendy A Wells, Keith D Paulsen, Brian W Pogue

Open access · hybridAbstract read
In one paragraph

Article in Physics in medicine and biology, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 12 papers.

0numbers the graph read from it
0cells of the map it votes in
12citing papers in PubMed
2.8field-weighted citation impact, top 10% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

12 citing papers in PubMed, 24 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
  9. Article
  10. Review
  11. Review of structured light in diffuse optical imaging.Journal of biomedical optics · 2018
    Review
  12. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 4 institutions in 1 country.

David M McClatchyThayer School of Engineering, Dartmouth College, 14 Engineering Dr., Hanover, NH 03755, United States of America.
Elizabeth J Rizzo
Jeff Meganck
Josh Kempner
Jared Vicory
Wendy A Wells
Keith D Paulsen
Brian W Pogue
Dartmouth–Hitchcock Medical Center · USDartmouth College · USPerkinElmer (United States) · USKitware (United States) · US

Funding

Optical Scatter Imaging System for Surgical Specimen Margin Assessment during Breast Conserving SurgeryR01CA192803 · NCI · DARTMOUTH COLLEGE · PI PAULSEN, KEITH D., POGUE, BRIAN WILLIAM · 2015 to 2019
$2.7M
Structured Light Scatteroscopy for Intraoperative Breast Tumor Margin AssessmentF31CA196308 · NCI · DARTMOUTH COLLEGE · PI MCCLATCHY III, DAVID · 2016 to 2018
$94k
NCI NIH HHS F31 CA196308NCI NIH HHS R01 CA192803
6 · The paper itself

Abstract

A multimodal micro-computed tomography (CT) and multi-spectral structured light imaging (SLI) system is introduced and systematically analyzed to test its feasibility to aid in margin delineation during breast conserving surgery (BCS). Phantom analysis of the micro-CT yielded a signal-to-noise ratio of 34, a contrast of 1.64, and a minimum detectable resolution of 240 μm for a 1.2 min scan. The SLI system, spanning wavelengths 490 nm to 800 nm and spatial frequencies up to 1.37 [Formula: see text], was evaluated with aqueous tissue simulating phantoms having variations in particle size distribution, scatter density, and blood volume fraction. The reduced scattering coefficient, [Formula: see text] and phase function parameter, γ, were accurately recovered over all wavelengths independent of blood volume fractions from 0% to 4%, assuming a flat sample geometry perpendicular to the imaging plane. The resolution of the optical system was tested with a step phantom, from which the modulation transfer function was calculated yielding a maximum resolution of 3.78 cycles per mm. The three dimensional spatial co-registration between the CT and optical imaging space was tested and shown to be accurate within 0.7 mm. A freshly resected breast specimen, with lobular carcinoma, fibrocystic disease, and adipose, was imaged with the system. The micro-CT provided visualization of the tumor mass and its spiculations, and SLI yielded superficial quantification of light scattering parameters for the malignant and benign tissue types. These results appear to be the first demonstration of SLI combined with standard medical tomography for imaging excised tumor specimens. While further investigations are needed to determine and test the spectral, spatial, and CT features required to classify tissue, this study demonstrates the ability of multimodal CT/SLI to quantify, visualize, and spatially navigate breast tumor specimens, which could potentially aid in the assessment of tumor margin status during BCS.

Indexed as

Image Processing, Computer-AssistedLightX-Ray MicrotomographyBreastBreast NeoplasmsCalibrationFemaleHumansMastectomy, SegmentalMultimodal ImagingPhantoms, ImagingSignal-To-Noise RatioTomography, X-Ray Computed

Identifiers

PMID29048330
PMCPMC5729028
OpenAlexW2766620092

What OpenQuestion holds

Textmetadata
LicenceTDM
Read underepoch 390

Registered trials

None linked