ArticleMedicine2017

In vitro biomechanical comparison after fixed- and mobile-core artificial cervical disc replacement versus fusion.

Jigang Lou, Yuanchao Li, Beiyu Wang, Yang Meng, Tingkui Wu, Hao Liu

Open access · goldAbstract readComparative Study
In one paragraph

Article in Medicine, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 11 papers.

0numbers the graph read from it
0cells of the map it votes in
11citing papers in PubMed
1.2field-weighted citation impact, top 19% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

11 citing papers in PubMed, 25 citations in OpenAlex.

  1. Article
  2. Article
  3. Changes in the centre of rotation and the anterior bone loss of the vertebral body in Mobi-C artificial disc replacement segments after cervical hybrid surgery: a retrospective study.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2024
    Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
  9. Article
  10. Article
  11. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 3 institutions in 1 country.

Jigang LouDepartment of Orthopedics, West China Hospital, Sichuan University, Chengdu, Sichuan Department of Biomechanical Research Laboratory, Shanghai Jiao Tong University, Shanghai, China.
Yuanchao Li
Beiyu Wang
Yang Meng
Tingkui Wu
Hao Liu
Sichuan University · CNShanghai Jiao Tong University · CNWest China Hospital of Sichuan University · CN

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

In vitro biomechanical analysis after cervical disc replacement (CDR) with a novel artificial disc prosthesis (mobile core) was conducted and compared with the intact model, simulated fusion, and CDR with a fixed-core prosthesis. The purpose of this experimental study was to analyze the biomechanical changes after CDR with a novel prosthesis and the differences between fixed- and mobile-core prostheses.Six human cadaveric C2-C7 specimens were biomechanically tested sequentially in 4 different spinal models: intact specimens, simulated fusion, CDR with a fixed-core prosthesis (Discover, DePuy), and CDR with a mobile-core prosthesis (Pretic-I, Trauson). Moments up to 2 Nm with a 75 N follower load were applied in flexion-extension, left and right lateral bending, and left and right axial rotation. The total range of motion (ROM), segmental ROM, and adjacent intradiscal pressure (IDP) were calculated and analyzed in 4 different spinal models, as well as the differences between 2 disc prostheses.Compared with the intact specimens, the total ROM, segmental ROM, and IDP at the adjacent segments showed no significant difference after arthroplasty. Moreover, CDR with a mobile-core prosthesis presented a little higher values of target segment (C5/6) and total ROM than CDR with a fixed-core prosthesis (P > .05). Besides, the difference in IDP at C4/5 after CDR with 2 prostheses was without statistical significance in all the directions of motion. However, the IDP at C6/7 after CDR with a mobile-core prosthesis was lower than CDR with a fixed-core prosthesis in flexion, extension, and lateral bending, with significant difference (P < .05), but not under axial rotation.CDR with a novel prosthesis was effective to maintain the ROM at the target segment and did not affect the ROM and IDP at the adjacent segments. Moreover, CDR with a mobile-core prosthesis presented a little higher values of target segment and total ROM, but lower IDP at the inferior adjacent segment than CDR with a fixed-core prosthesis.

Indexed as

Biomechanical PhenomenaCervical VertebraeCadaverHumansMaterials TestingModels, AnatomicProsthesis DesignProsthesis FittingRange of Motion, ArticularTotal Disc Replacement