Evidence map›Paper›PMID 28956513›Full record

ArticleInterventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences2018

The Medina Embolic Device: Karolinska experience.

P Bhogal, P A Brouwer, L Yeo, M Svensson, M Söderman

Abstract read
In one paragraph

Article in Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 12 papers.

0numbers the graph read from it
0cells of the map it votes in
12citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

12 citing papers in PubMed.

  1. Article
  2. Review
  3. Review
  4. Observational
  5. Article
  6. Review
  7. Endovascular Treatment of Intracranial Aneurysms.Life (Basel, Switzerland) · 2021
    Review
  8. Article
  9. Review
  10. Article
  11. Article
  12. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

P Bhogal1 Department of Neuroradiology, Karolinska University Hospital and Karolinska Institutet, Stockholm, Sweden.ORCID http://orcid.org/0000-0002-5514-5237
P A Brouwer1 Department of Neuroradiology, Karolinska University Hospital and Karolinska Institutet, Stockholm, Sweden.
L Yeo1 Department of Neuroradiology, Karolinska University Hospital and Karolinska Institutet, Stockholm, Sweden.
M Svensson4 Department of Neurosurgery , 59562 Karolinska University Hospital and Karolinska Institutet, Stockholm , Sweden.
M Söderman1 Department of Neuroradiology, Karolinska University Hospital and Karolinska Institutet, Stockholm, Sweden.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background The aim of this study was to report our single centre experience with the Medina Embolic Device (MED). Methods We performed a retrospective analysis of prospectively collected data to identify all patients treated with the MED. A total of 14 aneurysms (non-consecutive), in 13 patients, were treated including one ruptured and one partially thrombosed aneurysm. Fundus diameter was ≥5 mm in all cases. We evaluated the angiographic appearances, the clinical status, complications, and the need for adjunctive devices or repeat treatments. Results Aneurysm location was cavernous internal carotid artery (ICA; n = 1), supraclinoid ICA ( n = 1), terminal ICA ( n = 2), anterior communicating artery (AComA; n = 4), A2-3 ( n = 1), M1-2 junction ( n = 1), posterior communicating artery (PComA; n = 1), superior cerebellar artery (SCA; n = 1), and basilar tip ( n = 2). The average aneurysm fundus size was 8.6 mm (range 7-10 mm) and average neck size 3.75 mm (range 1.9-6.9 mm). Immediate angiographic results were modified Raymond-Roy occlusion classification (mRRC) I n = 2, mRRC II n = 1, mRRC IIIa n = 2, mRRC IIIb n = 2, the remaining 7 aneurysms showed complete opacification. At follow-up angiography (mean 5 months) mRRC I n = 5, mRRC II n = 5, mRRC IIIa n = 3, and persistent filling was seen in 1 aneurysm. Overall, four patients had repeat treatment and one is pending further treatment. Of the aneurysms treated with more than one MED, 75% showed complete occlusion at 6-month follow up whereas only one aneurysm treated with a single device showed complete occlusion. Overall, three patients had temporary complications and there were no deaths. Conclusions The MED is an intra-saccular flow-diverting device with satisfactory angiographic results and an acceptable safety profile. Use of a single MED cannot be recommended and further longer term studies are needed prior to widespread clinical use.

Indexed as

Cerebral AngiographyAdultAgedAneurysm, RupturedEndovascular ProceduresEquipment DesignFemaleHumansIntracranial AneurysmMaleMiddle AgedRetrospective StudiesTreatment OutcomeAneurysmflow diverterintra-saccularMedina Embolic Device

Identifiers

PMID28956513
PMCPMC5772534

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked