Evidence map›Paper›PMID 28859253›Full record

ArticleInternational journal for numerical methods in biomedical engineering2018

Flow features and device-induced blood trauma in CF-VADs under a pulsatile blood flow condition: A CFD comparative study.

Zengsheng Chen, Sofen K Jena, Guruprasad A Giridharan, Steven C Koenig, Mark S Slaughter, Bartley P Griffith, Zhongjun J Wu

Open access · greenAbstract read
In one paragraph

Article in International journal for numerical methods in biomedical engineering, 2018. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 22 papers.

0numbers the graph read from it
0cells of the map it votes in
22citing papers in PubMed
3.4field-weighted citation impact, top 7% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

22 citing papers in PubMed, 77 citations in OpenAlex.

  1. Review
  2. Article
  3. Review
  4. Numerical Analyses of Blood Pumps Under Realistic Operating Conditions.ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2026
    Article
  5. Article
  6. Article
  7. Article
  8. Review
  9. Article
  10. Article
  11. Article
  12. Article
  13. Article
  14. Article
  15. Article
  16. Article
  17. Article
  18. Article
  19. Article
  20. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 2 institutions in 1 country.

Zengsheng ChenDepartment of Surgery, University of Maryland School of Medicine, Baltimore, MD, 21201, USA.ORCID 0000-0003-3716-3156
Sofen K JenaDepartment of Cardiovascular and Thoracic Surgery, School of Medicine, University of Louisville, Louisville, KY, 40202, USA.
Guruprasad A GiridharanDepartment of Bioengineering, Speed School of Engineering, University of Louisville, Louisville, KY, 40292, USA.
Steven C KoenigDepartment of Cardiovascular and Thoracic Surgery, School of Medicine, University of Louisville, Louisville, KY, 40202, USA.
Mark S SlaughterDepartment of Cardiovascular and Thoracic Surgery, School of Medicine, University of Louisville, Louisville, KY, 40202, USA.
Bartley P GriffithDepartment of Surgery, University of Maryland School of Medicine, Baltimore, MD, 21201, USA.
Zhongjun J WuDepartment of Surgery, University of Maryland School of Medicine, Baltimore, MD, 21201, USA.ORCID 0000-0003-0807-7195
University of Louisville · USUniversity of Maryland, Baltimore · US

Funding

Shear-Induced Hemostatic Dysfunction and Bleeding in CF-VAD PatientsR01HL124170 · NHLBI · UNIVERSITY OF MARYLAND BALTIMORE · PI GRIFFITH, BARTLEY P, WU, ZHONGJUN JON · 2015 to 2018
$2.0M
Linking Multiscale Dynamics of Shear-induced Blood Damage to CFD modelingR01HL088100 · NHLBI · UNIVERSITY OF MARYLAND BALTIMORE · PI WU, ZHONGJUN JON · 2008 to 2011
$1.7M
NHLBI NIH HHS R01 HL088100NHLBI NIH HHS R01 HL124170
6 · The paper itself

Abstract

In this study, the flow features and device-associated blood trauma in 4 clinical ventricular assist devices (VADs; 2 implantable axial VADs, 1 implantable centrifugal VAD, and 1 extracorporeal VAD) were computationally analyzed under clinically relevant pulsatile flow conditions. The 4 VADs were operated at fixed pump speed at a mean rate of 4.5 L/min. Mean pressure difference, wall shear stress, volume distribution of scalar shear stress (SSS), and shear-induced hemolysis index (HI) were derived from the flow field of each VAD and were compared. The computationally predicted mean pressure difference across the 3 implantable VADs was ~70 mmHg, and the extracorporeal VAD was ~345 mmHg, which matched well with their reported pressure-flow curves. The axial VADs had higher mean wall shear stress and SSS compared with the centrifugal VADs. However, the residence time of the centrifugal VADs was much longer compared with the axial VADs because of the large volume of the centrifugal VADs. The highest SSS was observed in one axial VAD, and the longest exposure time was observed in 1 centrifugal VAD. These 2 VADs generated the highest HI. The shear-induced HI varied as a function of flow rate within each cardiac cycle. At fixed pump speed, the HI was greatest at low flow rate due to longer exposure time to shear stress compared with at high flow rate. Subsequently, we hypothesize that to reduce the risk of blood trauma during VAD support, shear stress magnitude and exposure time need to be minimized.

Indexed as

Heart-Assist DevicesPulsatile FlowAlgorithmsEquipment DesignHeart FailureHemolysisHumansShear Strengthblood fluid dynamicsblood traumaexposure timescalar shear stressVAD

Identifiers

PMID28859253
PMCPMC5803368
OpenAlexW2751541740

What OpenQuestion holds

Textmetadata
LicenceTDM
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.