ArticleBiomedical engineering online2017
16-Channel biphasic current-mode programmable charge balanced neural stimulation.
Article in Biomedical engineering online, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed, 15 citations in OpenAlex.
- Optimized Conditions for Electrical Tissue Stimulation with Biphasic, Charge-Balanced Impulses.Bioengineering (Basel, Switzerland) · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors at 2 institutions in 2 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundNeural stimulation is an important method used to activate or inhibit action potentials of the neuronal anatomical targets found in the brain, central nerve and peripheral nerve. The neural stimulator system produces biphasic pulses that deliver balanced charge into tissue from single or multichannel electrodes. The timing and amplitude of these biphasic pulses are precisely controlled by the neural stimulator software or imbedded algorithms. Amplitude mismatch between the anodic current and cathodic current of the biphasic pulse will cause permanently damage for the neural tissues. The main goal of our circuit and layout design is to implement a 16-channel biphasic current mode programmable neural stimulator with calibration to minimize the current mismatch caused by inherent complementary metal oxide semiconductor (CMOS) manufacturing processes.
methodsThis paper presents a 16-channel constant current mode neural stimulator chip. Each channel consists of a 7-bit controllable current DAC used as sink and source current driver. To reduce the LSB quantization error and the current mismatch, an automatic calibration circuit and flow diagram is presented in this paper. There are two modes of operation of the stimulator chip-namely, stimulation mode and calibration mode. The chip also includes a digital interface used to control the stimulator parameters and calibration levels specific for each individual channel.
resultsThis stimulator Application Specific Integrated Circuit (ASIC) is designed and fabricated in a 0.18 μm High-Voltage CMOS technology that allows for ±20 V power supply. The full-scale stimulation current was designed to be at 1 mA per channel. The output current was shown to be constant throughout the timing cycles over a wide range of electrode load impedances. The calibration circuit was also designed to reduce the effect of CMOS process variation of the P-channel metal oxide semiconductor (PMOS) and N-channel metal oxide semiconductor (NMOS) devices that will result in charge delivery to have less than 0.13% error.
conclusionsA 16-channel integrated biphasic neural stimulator chip with calibration is presented in this paper. The stimulator circuit design was simulated and the chip layout was completed. The chip layout was verified using design rules check (DRC) and layout versus schematic (LVS) design check using computer aided design (CAD) software. The test results we presented show constant current stimulation with charge balance error within 0.13% least-significant-bit (LSB). This LSB error was consistent throughout a variety stimulation patterns and electrode load impedances.